Decision Summary

Upload-ready calibration protocol overview
Instrument P83s1
Session 66
Location LabGSFCmanip
Calibration person Christoph Waldauf
CF file C:/Blick-master_19/data/calresults/Pandora83s1/session66/Pandora83s1_CF_session66.txt
Software version p1-9
Calibration vector -
Reference day 20240619
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P83s1
Session 66
Events 135
Figures 9
Action time 30m 14s

Action Timeline

43 action intervals
Jul 18, 2024Jul 19, 202402:30 PM07:44 AMFieldCalibrationCreateReferenceStrayLightConvolutionFieldCalibration 3s abortedCreateReference 1m 2s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedStrayLight 7s finishedStrayLight 0s finishedStrayLight 0s abortedConvolution 1m 0s abortedConvolution 3m 54s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedCreateReference 1m 25s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 53s abortedCreateReference 16s abortedCreateReference 22s finishedCreateReference 0s abortedCreateReference 23s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 20s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 12s finishedCreateReference 0s abortedCreateReference 38s abortedCreateReference 51s finishedFieldCalibration 9m 1s finishedFieldCalibration 0s abortedFieldCalibration 7m 48s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finished

Duration by Step

FieldCalibration16m 52sCreateReference8m 22sConvolution4m 54sStrayLight7s

Action Summary

Action Events Runs Aborted Time
Convolution 17 6 1 4m 54s
CreateReference 46 23 9 8m 22s
FieldCalibration 51 11 2 16m 52s
No action 15 0 0 0s
StrayLight 6 3 1 7s

Session Figures

9 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora83s1_WavelengthCalibration_SlitFuncParam.jpeg - 117 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora83s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 124 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora83s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 74 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora83s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 82 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora83s1_CreateReference_AODEST.jpeg - 46 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora83s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 81 KB
Create reference SENSCORRFIN (SENS)

Create reference SHIFT (CLOS)

Pandora83s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 111 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora83s1_FieldCalibration_VC_O3.jpeg - 80 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora83s1_FieldCalibration_SCREFINT_O3.jpeg - 55 KB
FieldCalibration SCREFINT O3

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O