Decision Summary

Upload-ready calibration protocol overview
Instrument P30s1
Session 2
Location LabGSFC
Calibration person Manuel Gebetsberger
CF file C:/Blick_19_GIT/data/calresults/Pandora30s1/session2/Pandora30s1_CF_session2.txt
Software version 1.9.4 / 1.9.5
Calibration vector -
Reference day -
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P30s1
Session 2
Events 161
Figures 1
Action time 23m 53s

Action Timeline

28 action intervals
Jul 13, 202309:07 PM10:56 PMStrayLightSensitivityCalibrationConvolutionStrayLight 7m 24s finishedStrayLight 0s finishedStrayLight 0s abortedStrayLight 9m 31s abortedStrayLight 3s abortedStrayLight 26s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedSensitivityCalibration 1m 4s abortedSensitivityCalibration 13s abortedStrayLight 37s finishedStrayLight 0s finishedStrayLight 0s abortedSensitivityCalibration 7s abortedSensitivityCalibration 1m 28s finishedSensitivityCalibration 0s finishedConvolution 2m 59s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finished

Duration by Step

StrayLight18m 1sConvolution2m 59sSensitivityCalibration2m 53s

Action Summary

Action Events Runs Aborted Time
Convolution 15 5 0 2m 59s
No action 12 0 0 0s
SensitivityCalibration 57 5 3 2m 53s
StrayLight 77 18 4 18m 1s

Session Figures

1 figure from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora30s1_WavelengthCalibration_SlitFuncParam.jpeg - 119 KB
Slit function parameters

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O