Decision Summary

Upload-ready calibration protocol overview
Instrument P242s2
Session 3
Location Trollhaugen
Calibration person Hanna Rohringer
CF file Pandora242s2_CF_v1d20230207.txt
Software version 1.9.5
Calibration vector calV1
Reference day 20241218
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P242s2
Session 3
Events 320
Figures 4
Action time 2h 46m

Action Timeline

97 action intervals
Dec 23, 2025Dec 24, 2025Dec 24, 2025Jan 01, 2026Jan 02, 202610:26 AMCreateReferenceConvolutionStrayLightFieldCalibrationAssignCalFileCreateReference 1m 48s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 11s finishedCreateReference 0s abortedCreateReference 1m 38s abortedCreateReference 3m 1s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 5s abortedConvolution 15m 10s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedCreateReference 13m 43s abortedCreateReference 3m 54s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 39s abortedStrayLight 4m 27s finishedStrayLight 0s finishedStrayLight 0s finishedConvolution 8m 6s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedCreateReference 5s abortedCreateReference 28s abortedCreateReference 1m 38s finishedCreateReference 0s abortedCreateReference 1m 13s abortedCreateReference 4m 54s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 30s abortedConvolution 15m 27s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedCreateReference 34s abortedCreateReference 2m 40s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 5s abortedCreateReference 32s abortedCreateReference 12s finishedCreateReference 0s abortedCreateReference 1m 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 34s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 8m 8s abortedConvolution 1s abortedCreateReference 46s abortedCreateReference 7m 47s finishedCreateReference 6m 44s finishedConvolution 15m 56s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedFieldCalibration 0s staleFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s staleFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 40m 58s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 2m 46s finished

Duration by Step

CreateReference1h 3mConvolution54m 41sFieldCalibration40m 58sStrayLight4m 27sAssignCalFile2m 46s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 2m 46s
Convolution 62 21 1 54m 41s
CreateReference 101 48 23 1h 3m
FieldCalibration 131 22 0 40m 58s
No action 18 0 0 0s
StrayLight 5 3 0 4m 27s

Session Figures

4 figures from Google Drive/figures
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora242s2_CreateReference_L1CHARFILT_OPEN.jpeg - 98 KB
Create reference L1CHARFILT

Create reference AODEST

Pandora242s2_CreateReference_AODEST.jpeg - 46 KB
Create reference AODEST
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=5.333e-03; wrmsmax=2.847e-01; atvarmax(%)=5.000e+00; amflim=(1.500,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=220; timeaver(min)=0.00; redu=2; scref=(1.766e-04,1.699e-04,1.699e-04); uscref=(1.180e-05,2.156e-06,1.208e-05)

FieldCalibration MLE NO2

Pandora242s2_FieldCalibration_MLE_NO2.jpeg - 153 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora242s2_FieldCalibration_VCMLE_NO2.jpeg - 156 KB
FieldCalibration VCMLE NO2

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O