Decision Summary

Upload-ready calibration protocol overview
Instrument P240s1
Session 16
Location Thessaloniki
Calibration person Ragi Rajagopalan
CF file C:/Blick-v1.9/data/calresults/Pandora240s1/session16/Pandora240s1_CF_session16.txt
Software version 1.9.6
Calibration vector -
Reference day 20230713
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P240s1
Session 16
Events 315
Figures 17
Action time 1h 52m

Action Timeline

108 action intervals
Sep 23, 2024Sep 24, 202408:22 AM09:53 AMCreateReferenceWavelengthCalibrationStrayLightConvolutionFieldCalibrationCreateReference 54s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 20s finishedCreateReference 0s abortedCreateReference 2m 46s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 2s abortedCreateReference 18s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 21s abortedCreateReference 16s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 5s abortedCreateReference 37s abortedCreateReference 48s abortedCreateReference 52s abortedCreateReference 28s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 40s abortedCreateReference 36s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 27s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 23s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 24s abortedCreateReference 1m 44s abortedWavelengthCalibration 11s finishedWavelengthCalibration 1h 0m abortedWavelengthCalibration 41s abortedWavelengthCalibration 17s finishedCreateReference 2m 55s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 48s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 21s abortedCreateReference 5s abortedCreateReference 23s abortedCreateReference 11s abortedStrayLight 7s abortedStrayLight 8s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedCreateReference 11m 3s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedStrayLight 7s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 0s finishedStrayLight 27s finishedStrayLight 0s finishedStrayLight 0s finishedConvolution 3m 10s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedCreateReference 8s finishedCreateReference 0s abortedCreateReference 54s abortedCreateReference 1m 18s abortedCreateReference 1m 18s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 3m 7s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 2m 23s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 5m 18s finishedCreateReference 0s finishedFieldCalibration 2s aborted

Duration by Step

WavelengthCalibration1h 2mCreateReference46m 10sConvolution3m 10sStrayLight50sFieldCalibration2s

Action Summary

Action Events Runs Aborted Time
Convolution 15 5 0 3m 10s
CreateReference 145 71 31 46m 10s
FieldCalibration 2 1 1 2s
No action 11 0 0 0s
StrayLight 131 27 1 50s
WavelengthCalibration 11 4 2 1h 2m

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora240s1_WavelengthCalibration_SlitFuncParam.jpeg - 227 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora240s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 137 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora240s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 117 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora240s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 160 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora240s1_CreateReference_AODEST.jpeg - 66 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora240s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 169 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora240s1_CreateReference_DISPFIT(DISP).jpeg - 112 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora240s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 201 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora240s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 114 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora240s1_FieldCalibration_VC_O3.jpeg - 100 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora240s1_FieldCalibration_SCREFINT_O3.jpeg - 79 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora240s1_FieldCalibration_MLE_NO2.jpeg - 173 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora240s1_FieldCalibration_VCMLE_NO2.jpeg - 136 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora240s1_FieldCalibration_MLE_SO2.jpeg - 83 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora240s1_FieldCalibration_VCMLE_SO2.jpeg - 77 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora240s1_FieldCalibration_MLE_HCHO.jpeg - 89 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora240s1_FieldCalibration_VCMLE_HCHO.jpeg - 129 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O