Decision Summary

Upload-ready calibration protocol overview
Instrument P234s1
Session 78
Location Izana
Calibration person Manuel Gebetsberger
CF file C:/CalibrationAnalysisOperational/Blick-1.9.5/data/calresults/Pandora234s1/session78/Pandora234s1_CF_session78.txt
Software version 1.9.5
Calibration vector -
Reference day -
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P234s1
Session 78
Events 64
Figures 3
Action time 2h 32m

Action Timeline

19 action intervals
Jun 24, 2024 - Jun 25, 202409:18 PM09:05 AMConvolutionStrayLightConvolution 30s abortedStrayLight 9s finishedStrayLight 0s finishedStrayLight 0s abortedStrayLight 40s finishedStrayLight 0s finishedStrayLight 0s finishedConvolution 35m 55s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 37s abortedConvolution 34m 58s finishedConvolution 0s finishedConvolution 0s abortedConvolution 1h 19m finishedConvolution 0s finishedConvolution 0s finished

Duration by Step

Convolution2h 31mStrayLight49s

Action Summary

Action Events Runs Aborted Time
Convolution 36 13 3 2h 31m
No action 17 0 0 0s
StrayLight 11 6 1 49s

Session Figures

3 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora234s1_WavelengthCalibration_SlitFuncParam.jpeg - 116 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora234s1_CreateReference_L1CHARFILT_OPEN.jpeg - 91 KB
Create reference L1CHARFILT

Create reference AODEST

Pandora234s1_CreateReference_AODEST.jpeg - 48 KB
Create reference AODEST

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O