Decision Summary

Upload-ready calibration protocol overview
Instrument P182s1
Session 11
Location Tel-Aviv
Calibration person Christoph Waldauf
CF file D:/Blick-master_19/data/calresults/Pandora182s1/session11/Pandora182s1_CF_session11.txt
Software version p1-9
Calibration vector -
Reference day 20250612
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P182s1
Session 11
Events 127
Figures 9
Action time 47m 7s

Action Timeline

44 action intervals
Sep 19, 202509:56 AM11:59 AMCreateReferenceFieldCalibrationCreateReference 6m 41s finishedCreateReference 0s abortedCreateReference 33s finishedCreateReference 0s abortedCreateReference 21s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 39s finishedCreateReference 0s abortedCreateReference 48s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 52s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 22m 9s finishedCreateReference 0s abortedCreateReference 6s finishedCreateReference 0s abortedCreateReference 3m 18s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 36s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 36s finishedCreateReference 0s finishedFieldCalibration 2m 4s abortedFieldCalibration 6m 14s finishedFieldCalibration 0s abortedFieldCalibration 41s finishedFieldCalibration 0s abortedFieldCalibration 1m 30s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finished

Duration by Step

CreateReference36m 39sFieldCalibration10m 29s

Action Summary

Action Events Runs Aborted Time
CreateReference 60 30 10 36m 39s
FieldCalibration 56 14 3 10m 29s
No action 11 0 0 0s

Session Figures

9 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora182s1_WavelengthCalibration_SlitFuncParam.jpeg - 117 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora182s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 108 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora182s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 92 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora182s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 92 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora182s1_CreateReference_AODEST.jpeg - 45 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora182s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 75 KB
Create reference SENSCORRFIN (SENS)

Create reference SHIFT (CLOS)

Pandora182s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 93 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora182s1_FieldCalibration_VC_O3.jpeg - 63 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora182s1_FieldCalibration_SCREFINT_O3.jpeg - 52 KB
FieldCalibration SCREFINT O3

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O