Decision Summary

Upload-ready calibration protocol overview
Instrument P138s1
Session 5
Location Rome-SAP
Calibration person Markus Kilian
CF file Pandora138s1_CF_v5d20190827.txt
Software version 1.9.5
Calibration vector calV1
Reference day -
Overall status development
O3Ext: YesO3Synt: YesNO2: YesSO2: YesHCHO: YesH2O: Yes
Instrument P138s1
Session 5
Events 139
Figures 6
Action time 1h 8m

Action Timeline

36 action intervals
Oct 13, 2025Oct 14, 202501:26 PM06:07 AMFieldCalibrationAssignCalFileFieldCalibration 33m 30s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 23m 57s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 3m 12s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 21s abortedFieldCalibration 3m 20s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 3m 20s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 35s finished

Duration by Step

FieldCalibration1h 7mAssignCalFile35s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 35s
FieldCalibration 127 35 1 1h 7m
No action 9 0 0 0s

Session Figures

6 figures from Google Drive/figures
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=5.463e-04; wrmsmax=9.000e-04; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=43; timeaver(min)=0.00; redu=2; scref=(1.370e-04,1.314e-04,1.258e-04); uscref=(6.975e-06,6.698e-06,9.433e-06)

FieldCalibration MLE NO2

Pandora138s1_FieldCalibration_MLE_NO2.jpeg - 110 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora138s1_FieldCalibration_VCMLE_NO2.jpeg - 76 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=8.053e-03; wrmsmax=9.000e-04; atvarmax(%)=4.000e+01; amflim=(1.000,2.500); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(6.419e-05); uscref=(3.448e-05)

FieldCalibration MLE SO2

Pandora138s1_FieldCalibration_MLE_SO2.jpeg - 78 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora138s1_FieldCalibration_VCMLE_SO2.jpeg - 80 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=6.356e-03; wrmsmax=9.000e-04; atvarmax(%)=4.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=12; timeaver(min)=0.00; redu=2; scref=(1.928e-04); uscref=(2.248e-05)

FieldCalibration MLE HCHO

Pandora138s1_FieldCalibration_MLE_HCHO.jpeg - 76 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora138s1_FieldCalibration_VCMLE_HCHO.jpeg - 84 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
O3 products look rather good

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
EMLE used; looks good

SO2

Decision for SO2
Yes
daily cycles okay, but rather high uncertainty

HCHO

Decision for HCHO
Yes
EMLE used

H2O

Decision for H2O
Yes