Decision Summary

Upload-ready calibration protocol overview
Instrument P120s1
Session 20
Location LabLuftBlick
Calibration person Christoph Waldauf
CF file D:/Blick-master_19/data/calresults/Pandora120s1/session20/Pandora120s1_CF_session20.txt
Software version 1.9.5
Calibration vector -
Reference day -
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P120s1
Session 20
Events 208
Figures 1
Action time 45m 48s

Action Timeline

24 action intervals
Sep 29, 202502:42 PM04:18 PMCheckFiltersCheckNoiseWaveAnalyzeDarkAnalyzeBrightWavelengthCalibrationNDFiltersStrayLightSensitivityCalibrationConvolutionCheckFilters 35s finishedCheckNoiseWave 23s finishedAnalyzeDark 1m 16s finishedAnalyzeBright 13m 51s finishedWavelengthCalibration 40s abortedWavelengthCalibration 1m 39s abortedWavelengthCalibration 7m 46s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedNDFilters 4m 4s finishedStrayLight 5s finishedStrayLight 0s finishedStrayLight 0s abortedSensitivityCalibration 10m 20s finishedSensitivityCalibration 0s abortedSensitivityCalibration 4m 12s finishedSensitivityCalibration 0s finishedConvolution 56s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finished

Duration by Step

SensitivityCalibration14m 32sAnalyzeBright13m 51sWavelengthCalibration10m 5sNDFilters4m 4sAnalyzeDark1m 16sConvolution56sCheckFilters35sCheckNoiseWave23sStrayLight5s

Action Summary

Action Events Runs Aborted Time
AnalyzeBright 14 1 0 13m 51s
AnalyzeDark 11 1 0 1m 16s
CheckFilters 4 1 0 35s
CheckNoiseWave 6 1 0 23s
Convolution 16 5 0 56s
NDFilters 28 1 0 4m 4s
No action 9 0 0 0s
SensitivityCalibration 90 4 1 14m 32s
StrayLight 5 3 1 5s
WavelengthCalibration 25 7 2 10m 5s

Session Figures

1 figure from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora120s1_WavelengthCalibration_SlitFuncParam.jpeg - 116 KB
Slit function parameters

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O