Decision Summary

Upload-ready calibration protocol overview
Instrument P117s1
Session 2
Location Rome-SAP
Calibration person Markus Kilian
CF file Pandora117s1_CF_v2d20250721.txt
Software version 1.9.5
Calibration vector calV1
Reference day -
Overall status development
O3Ext: YesO3Synt: YesNO2: YesSO2: YesHCHO: YesH2O: Yes
Instrument P117s1
Session 2
Events 130
Figures 15
Action time 48m 37s

Action Timeline

33 action intervals
Oct 10, 202506:50 AM08:35 AMFieldCalibrationCreateReferenceAssignCalFileFieldCalibration 2s abortedCreateReference 32s abortedFieldCalibration 23m 38s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 15m 9s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 5m 41s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 2m 54s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 41s finished

Duration by Step

FieldCalibration47m 24sAssignCalFile41sCreateReference32s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 41s
CreateReference 2 1 1 32s
FieldCalibration 116 31 1 47m 24s
No action 9 0 0 0s

Session Figures

15 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora117s1_WavelengthCalibration_SlitFuncParam.jpeg - 118 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora117s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 116 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora117s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 74 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora117s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 91 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora117s1_CreateReference_AODEST.jpeg - 45 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora117s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 76 KB
Create reference SENSCORRFIN (SENS)

Create reference SHIFT (CLOS)

Pandora117s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 87 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora117s1_FieldCalibration_VC_O3.jpeg - 61 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora117s1_FieldCalibration_SCREFINT_O3.jpeg - 52 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=4.034e-05; wrmsmax=9.000e-04; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=30; timeaver(min)=0.00; redu=2; scref=(1.536e-04,1.403e-04,1.413e-04); uscref=(9.211e-06,5.901e-06,1.129e-05)

FieldCalibration MLE NO2

Pandora117s1_FieldCalibration_MLE_NO2.jpeg - 113 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora117s1_FieldCalibration_VCMLE_NO2.jpeg - 80 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=2.096e-02; wrmsmax=9.000e-04; atvarmax(%)=4.000e+01; amflim=(1.000,2.500); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(4.146e-05); uscref=(1.343e-05)

FieldCalibration MLE SO2

Pandora117s1_FieldCalibration_MLE_SO2.jpeg - 66 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora117s1_FieldCalibration_VCMLE_SO2.jpeg - 59 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=2.625e-02; wrmsmax=9.000e-04; atvarmax(%)=4.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=2; scref=(2.968e-04); uscref=(1.979e-05)

FieldCalibration MLE HCHO

Pandora117s1_FieldCalibration_MLE_HCHO.jpeg - 67 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora117s1_FieldCalibration_VCMLE_HCHO.jpeg - 66 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
O3 products okay

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
looks good, E-MLE used

SO2

Decision for SO2
Yes
looks good

HCHO

Decision for HCHO
Yes
looks good, E-MLE used

H2O

Decision for H2O
Yes