Decision Summary

Upload-ready calibration protocol overview
Instrument P117s1
Session 15
Location Rome-SAP
Calibration person Markus Kilian
CF file Pandora117s1_CF_v15d20180809.txt
Software version 1.9.5
Calibration vector -
Reference day 20180531
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P117s1
Session 15
Events 333
Figures 15
Action time 2h 13m

Action Timeline

96 action intervals
Mar 03, 202608:42 AM11:59 AMCreateReferenceConvolutionFieldCalibrationAssignCalFileCreateReference 5m 16s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 12s finishedCreateReference 0s abortedCreateReference 49s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 50s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 19s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 6s abortedCreateReference 18s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 19s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 47s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedConvolution 17m 9s finishedConvolution 0s finishedConvolution 0s finishedCreateReference 22s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 20s abortedCreateReference 6s finishedCreateReference 0s abortedCreateReference 38s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 20s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 12s finishedCreateReference 0s finishedCreateReference 1s abortedFieldCalibration 2m 32s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 15m 46s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 18m 29s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 6s abortedFieldCalibration 20s abortedFieldCalibration 13s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 5m 57s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 31m 56s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 30m 32s finished

Duration by Step

FieldCalibration1h 15mAssignCalFile30m 32sConvolution17m 9sCreateReference10m 55s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 30m 32s
Convolution 9 3 0 17m 9s
CreateReference 102 48 15 10m 55s
FieldCalibration 207 44 3 1h 15m
No action 12 0 0 0s

Session Figures

15 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora117s1_WavelengthCalibration_SlitFuncParam.jpeg - 118 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora117s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 128 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora117s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 100 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora117s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 96 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora117s1_CreateReference_AODEST.jpeg - 45 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora117s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 83 KB
Create reference SENSCORRFIN (SENS)

Create reference SHIFT (CLOS)

Pandora117s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 90 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora117s1_FieldCalibration_VC_O3.jpeg - 79 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora117s1_FieldCalibration_SCREFINT_O3.jpeg - 53 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=8.847e-03; wrmsmax=9.000e-04; atvarmax(%)=6.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=2; scref=(3.476e-04,3.148e-04,3.328e-04); uscref=(1.439e-05,1.581e-05,1.864e-05)

FieldCalibration MLE NO2

Pandora117s1_FieldCalibration_MLE_NO2.jpeg - 93 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora117s1_FieldCalibration_VCMLE_NO2.jpeg - 87 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=2.264e-02; wrmsmax=9.000e-04; atvarmax(%)=6.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(7.320e-05); uscref=(1.928e-05)

FieldCalibration MLE SO2

Pandora117s1_FieldCalibration_MLE_SO2.jpeg - 67 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora117s1_FieldCalibration_VCMLE_SO2.jpeg - 73 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=5.312e-02; wrmsmax=9.000e-04; atvarmax(%)=6.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=30; timeaver(min)=0.00; redu=2; scref=(2.479e-04); uscref=(1.012e-05)

FieldCalibration MLE HCHO

Pandora117s1_FieldCalibration_MLE_HCHO.jpeg - 69 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora117s1_FieldCalibration_VCMLE_HCHO.jpeg - 100 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O