Decision Summary

Upload-ready calibration protocol overview
Instrument P117s1
Session 12
Location Rome-SAP
Calibration person Christoph Waldauf
CF file Pandora117s1_CF_v12d20200604.txt
Software version 1.9.5
Calibration vector calV2
Reference day 20210622
Overall status development
O3Ext: YesO3Synt: YesNO2: With CareSO2: YesHCHO: YesH2O: Not answered
Instrument P117s1
Session 12
Events 328
Figures 15
Action time 6h 12m

Action Timeline

70 action intervals
Jan 19, 2026Jan 20, 202609:43 AMCreateReferenceFieldCalibrationAssignCalFileCreateReference 12m 5s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 1s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 3m 14s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 8s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedFieldCalibration 3s abortedCreateReference 49s abortedCreateReference 6m 28s abortedCreateReference 55s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 27s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 42s finishedCreateReference 0s finishedFieldCalibration 6m 20s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 10s abortedFieldCalibration 4h 27m finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 39m 9s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 8m 56s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 21m 48s finishedFieldCalibration 0s abortedCreateReference 2s abortedFieldCalibration 12s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 1m 6s finished

Duration by Step

FieldCalibration5h 44mCreateReference26m 50sAssignCalFile1m 6s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 1m 6s
CreateReference 57 27 9 26m 50s
FieldCalibration 259 42 3 5h 44m
No action 9 0 0 0s

Session Figures

15 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora117s1_WavelengthCalibration_SlitFuncParam.jpeg - 143 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora117s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 102 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora117s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 73 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora117s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 94 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora117s1_CreateReference_AODEST.jpeg - 49 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora117s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 74 KB
Create reference SENSCORRFIN (SENS)

Create reference SHIFT (CLOS)

Pandora117s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 108 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora117s1_FieldCalibration_VC_O3.jpeg - 64 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora117s1_FieldCalibration_SCREFINT_O3.jpeg - 52 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.438e-03; wrmsmax=2.824e+00; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=30; timeaver(min)=0.00; redu=2; scref=(1.410e-04,1.286e-04,1.280e-04); uscref=(8.538e-06,5.034e-06,1.154e-05)

FieldCalibration MLE NO2

Pandora117s1_FieldCalibration_MLE_NO2.jpeg - 112 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora117s1_FieldCalibration_VCMLE_NO2.jpeg - 63 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.062e-02; wrmsmax=3.397e-02; atvarmax(%)=2.000e+01; amflim=(1.000,2.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(5.444e-05); uscref=(1.993e-05)

FieldCalibration MLE SO2

Pandora117s1_FieldCalibration_MLE_SO2.jpeg - 67 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora117s1_FieldCalibration_VCMLE_SO2.jpeg - 62 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=3.152e-04; wrmsmax=4.033e-03; atvarmax(%)=2.000e+01; amflim=(1.000,2.500); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=2; scref=(2.696e-04); uscref=(2.015e-05)

FieldCalibration MLE HCHO

Pandora117s1_FieldCalibration_MLE_HCHO.jpeg - 67 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora117s1_FieldCalibration_VCMLE_HCHO.jpeg - 97 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
looks good, interpolation used in AXC. O3Temp slightly lower than in earlier VP

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
With Care
MLE used, looks good

SO2

Decision for SO2
Yes
looks good

HCHO

Decision for HCHO
Yes
E-MLE used, looks good

H2O

Decision for H2O