Decision Summary
Upload-ready calibration protocol overview
Instrument
P117s1
Session
12
Location
Rome-SAP
Calibration person
Christoph Waldauf
CF file
Pandora117s1_CF_v12d20200604.txt
Software version
1.9.5
Calibration vector
calV2
Reference day
20210622
Overall status
development
O3Ext: YesO3Synt: YesNO2: With CareSO2: YesHCHO: YesH2O: Not answered
Action Timeline
70 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 3 | 1 | 0 | 1m 6s |
| CreateReference | 57 | 27 | 9 | 26m 50s |
| FieldCalibration | 259 | 42 | 3 | 5h 44m |
| No action | 9 | 0 | 0 | 0s |
Session Figures
15 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora117s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 102 KBCreate reference SELSPECS
Pandora117s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 73 KBCreate reference SLERR (STRAY)
Pandora117s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 94 KBCreate reference AODEST
Pandora117s1_CreateReference_AODEST.jpeg - 49 KBCreate reference SENSCORRFIN (SENS)
Pandora117s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 74 KBCreate reference SHIFT (CLOS)
Pandora117s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 108 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
looks good, interpolation used in AXC. O3Temp slightly lower than in earlier VP
O3Synt
Decision for O3Synt
Yes
NO2
Decision for NO2
With Care
MLE used, looks good
SO2
Decision for SO2
Yes
looks good
HCHO
Decision for HCHO
Yes
E-MLE used, looks good
H2O
Decision for H2O