Decision Summary
Upload-ready calibration protocol overview
Instrument
P117s1
Session
11
Location
Rome-SAP
Calibration person
Christoph Waldauf
CF file
Pandora117s1_CF_v11d20170130.txt
Software version
1.9.5
Calibration vector
calV1
Reference day
20190613
Overall status
development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: With CareHCHO: With CareH2O: Unusable
Action Timeline
257 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeDark | 2 | 1 | 1 | 3s |
| AssignCalFile | 3 | 1 | 0 | 3m 10s |
| CheckFilters | 52 | 16 | 10 | 13m 29s |
| Convolution | 110 | 38 | 4 | 46m 55s |
| CreateReference | 192 | 92 | 37 | 1h 9m |
| FieldCalibration | 260 | 42 | 6 | 1h 57m |
| Manual action | 1 | 0 | 0 | 0s |
| NDFilters | 87 | 7 | 6 | 6m 14s |
| No action | 23 | 0 | 0 | 0s |
| SensitivityCalibration | 552 | 41 | 25 | 50m 12s |
| StrayLight | 10 | 6 | 2 | 1m 2s |
| WavelengthCalibration | 36 | 12 | 6 | 8m 0s |
Session Figures
15 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora117s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 103 KBCreate reference SELSPECS
Pandora117s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 74 KBCreate reference SLERR (STRAY)
Pandora117s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 94 KBCreate reference AODEST
Pandora117s1_CreateReference_AODEST.jpeg - 46 KBCreate reference SENSCORRFIN (SENS)
Pandora117s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 75 KBCreate reference SHIFT (CLOS)
Pandora117s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 106 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
AXC looked good, no extrapol.; the temp looks good in the beginning, in the later timeperiod it gets a small inverse-u shape; also the wrms has the weird asymmetrical shape with a 'dip' every forenoon; ous looks good in general, also clear wrms shape; out looks okay, wrms seems high but still okay
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
E-MLE used
SO2
Decision for SO2
With Care
data looks okay, around 0 and no clear shapes, but in winterperiod partly flagged due to wrms, but it comes back; what looks strange is the shape of the wrms (asymmetrical daily repeating shape)
HCHO
Decision for HCHO
With Care
E-MLE used, during winterperiod a lot flagged due to wrms, but it comes back in march; data looks okay
H2O
Decision for H2O
Unusable
clear inverse-u shape, tried a lower value but it still has the shape