Decision Summary
Upload-ready calibration protocol overview
Instrument
P115s1
Session
14
Location
Rome-ISAC
Calibration person
Markus Kilian
CF file
Pandora115s1_CF_v14d20240719.txt
Software version
1.9.5
Calibration vector
calV1
Reference day
-
Overall status
development
O3Ext: YesO3Synt: YesNO2: YesSO2: With CareHCHO: YesH2O: Yes
Action Timeline
44 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 3 | 1 | 0 | 4m 2s |
| CreateReference | 2 | 1 | 1 | 10s |
| FOVDetermination | 2 | 1 | 1 | 2s |
| FieldCalibration | 184 | 41 | 3 | 1h 2m |
| No action | 12 | 0 | 0 | 0s |
Session Figures
15 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora115s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 108 KBCreate reference SELSPECS
Pandora115s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 73 KBCreate reference SLERR (STRAY)
Pandora115s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 91 KBCreate reference AODEST
Pandora115s1_CreateReference_AODEST.jpeg - 45 KBCreate reference SENSCORRFIN (SENS)
Pandora115s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 81 KBCreate reference SHIFT (CLOS)
Pandora115s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 117 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
O3Ext has inverse u-shape on some days
O3Synt
Decision for O3Synt
Yes
NO2
Decision for NO2
Yes
looks good, E-MLE used
SO2
Decision for SO2
With Care
wiggly shape, some days V-shape
HCHO
Decision for HCHO
Yes
looks good, E-MLE used, flagged until Dec 2024, cause of instrumental problem
H2O
Decision for H2O
Yes