Decision Summary

Upload-ready calibration protocol overview
Instrument P115s1
Session 14
Location Rome-ISAC
Calibration person Markus Kilian
CF file Pandora115s1_CF_v14d20240719.txt
Software version 1.9.5
Calibration vector calV1
Reference day -
Overall status development
O3Ext: YesO3Synt: YesNO2: YesSO2: With CareHCHO: YesH2O: Yes
Instrument P115s1
Session 14
Events 203
Figures 15
Action time 1h 6m

Action Timeline

44 action intervals
Oct 09, 202509:24 AM12:09 PMFOVDeterminationCreateReferenceFieldCalibrationAssignCalFileFOVDetermination 2s abortedCreateReference 10s abortedFieldCalibration 25m 51s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 6s finishedFieldCalibration 51s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 24m 23s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 4m 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 3m 22s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 14s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 3m 45s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 5s abortedAssignCalFile 4m 2s finished

Duration by Step

FieldCalibration1h 2mAssignCalFile4m 2sCreateReference10sFOVDetermination2s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 4m 2s
CreateReference 2 1 1 10s
FOVDetermination 2 1 1 2s
FieldCalibration 184 41 3 1h 2m
No action 12 0 0 0s

Session Figures

15 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora115s1_WavelengthCalibration_SlitFuncParam.jpeg - 124 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora115s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 108 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora115s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 73 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora115s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 91 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora115s1_CreateReference_AODEST.jpeg - 45 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora115s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 81 KB
Create reference SENSCORRFIN (SENS)

Create reference SHIFT (CLOS)

Pandora115s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 117 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora115s1_FieldCalibration_VC_O3.jpeg - 62 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora115s1_FieldCalibration_SCREFINT_O3.jpeg - 53 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=9.291e-05; wrmsmax=9.000e-04; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=39; timeaver(min)=0.00; redu=2; scref=(1.491e-04,1.265e-04,1.331e-04); uscref=(1.056e-05,6.544e-06,1.106e-05)

FieldCalibration MLE NO2

Pandora115s1_FieldCalibration_MLE_NO2.jpeg - 113 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora115s1_FieldCalibration_VCMLE_NO2.jpeg - 75 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.013e-02; wrmsmax=9.000e-04; atvarmax(%)=4.000e+01; amflim=(1.000,2.500); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(6.276e-05); uscref=(4.779e-05)

FieldCalibration MLE SO2

Pandora115s1_FieldCalibration_MLE_SO2.jpeg - 68 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora115s1_FieldCalibration_VCMLE_SO2.jpeg - 59 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=4.564e-03; wrmsmax=9.000e-04; atvarmax(%)=4.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=2; scref=(1.689e-04); uscref=(1.181e-05)

FieldCalibration MLE HCHO

Pandora115s1_FieldCalibration_MLE_HCHO.jpeg - 77 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora115s1_FieldCalibration_VCMLE_HCHO.jpeg - 83 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
O3Ext has inverse u-shape on some days

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
looks good, E-MLE used

SO2

Decision for SO2
With Care
wiggly shape, some days V-shape

HCHO

Decision for HCHO
Yes
looks good, E-MLE used, flagged until Dec 2024, cause of instrumental problem

H2O

Decision for H2O
Yes