Decision Summary

Upload-ready calibration protocol overview
Instrument P105s1
Session 12
Location Helsinki
Calibration person Christoph Waldauf
CF file D:/Blick-master_19/data/calresults/Pandora105s1/session12/Pandora105s1_CF_session12.txt
Software version p1-9
Calibration vector -
Reference day 20250610
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P105s1
Session 12
Events 157
Figures 8
Action time 32m 32s

Action Timeline

45 action intervals
Oct 07, 202512:50 PM03:30 PMCreateReferenceSensitivityCalibrationConvolutionFieldCalibrationCreateReference 1m 15s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedSensitivityCalibration 21s finishedSensitivityCalibration 0s finishedConvolution 53s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedCreateReference 37s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 34s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 18s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 27s abortedCreateReference 1m 5s abortedCreateReference 35s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 37s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 40s finishedCreateReference 0s finishedFieldCalibration 16m 30s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 6m 31s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 8s finishedFieldCalibration 0s finished

Duration by Step

FieldCalibration23m 10sCreateReference8m 8sConvolution53sSensitivityCalibration21s

Action Summary

Action Events Runs Aborted Time
Convolution 16 5 0 53s
CreateReference 53 26 8 8m 8s
FieldCalibration 43 12 1 23m 10s
No action 9 0 0 0s
SensitivityCalibration 36 2 0 21s

Session Figures

8 figures from Google Drive/figures
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora105s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 113 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora105s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 78 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora105s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 77 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora105s1_CreateReference_AODEST.jpeg - 47 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora105s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 85 KB
Create reference SENSCORRFIN (SENS)

Create reference SHIFT (CLOS)

Pandora105s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 110 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora105s1_FieldCalibration_VC_O3.jpeg - 59 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora105s1_FieldCalibration_SCREFINT_O3.jpeg - 54 KB
FieldCalibration SCREFINT O3

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O