Decision Summary

Upload-ready calibration protocol overview
Instrument P94s1
Session 4
Location GuatemalaCity
Calibration person Leonie Haunold
CF file Pandora94s1_CF_v4d20250225.txt
Software version 1.8.92
Calibration vector calV1
Reference day 20260215
Overall status development
O3Ext: UnusableO3Synt: With CareNO2: YesSO2: YesHCHO: YesH2O: Yes
Instrument P94s1
Session 4
Events 382
Figures 17
Action time 5h 31m

Action Timeline

108 action intervals
Apr 22, 2026Apr 23, 2026Apr 29, 2026May 08, 202601:49 PMCreateReferenceWavelengthCalibrationFieldCalibrationAssignCalFileCreateReference 5s abortedCreateReference 42s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 14s finishedCreateReference 0s abortedCreateReference 1m 4s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 39s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedWavelengthCalibration 9s finishedCreateReference 54s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 3h 53m finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 19s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 19s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 14s abortedCreateReference 14s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 3m 11s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 25s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 31s finishedCreateReference 0s finishedFieldCalibration 2m 26s abortedFieldCalibration 2m 32s finishedFieldCalibration 0s abortedFieldCalibration 10s abortedFieldCalibration 13s abortedFieldCalibration 18s abortedFieldCalibration 32m 7s finishedFieldCalibration 23s abortedFieldCalibration 10m 49s finishedFieldCalibration 46s abortedFieldCalibration 12s abortedFieldCalibration 22s abortedFieldCalibration 2s abortedFieldCalibration 15s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s staleFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 5s abortedFieldCalibration 14s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 12s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 16m 41s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 12m 25s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 17s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 29s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 3s abortedAssignCalFile 54s finishedWavelengthCalibration 0s staleFieldCalibration 4s abortedAssignCalFile 5m 36s abortedWavelengthCalibration 0s open

Duration by Step

CreateReference4h 3mFieldCalibration1h 21mAssignCalFile6m 30sWavelengthCalibration9s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 5 2 1 6m 30s
CreateReference 89 44 13 4h 3m
FieldCalibration 257 58 17 1h 21m
No action 24 0 0 0s
WavelengthCalibration 7 1 0 9s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora94s1_WavelengthCalibration_SlitFuncParam.jpeg - 145 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora94s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 101 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora94s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 95 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora94s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 87 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora94s1_CreateReference_AODEST.jpeg - 37 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora94s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 84 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora94s1_CreateReference_DISPFIT(DISP).jpeg - 87 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora94s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 122 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora94s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 102 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora94s1_FieldCalibration_VC_O3.jpeg - 70 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora94s1_FieldCalibration_SCREFINT_O3.jpeg - 54 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=4.620e-04; wrmsmax=2.072e+00; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=100; timeaver(min)=0.00; redu=2; scref=(1.226e-04,1.043e-04,1.098e-04); uscref=(1.840e-05,1.621e-05,1.781e-05)

FieldCalibration MLE NO2

Pandora94s1_FieldCalibration_MLE_NO2.jpeg - 167 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora94s1_FieldCalibration_VCMLE_NO2.jpeg - 95 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=4.467e-03; wrmsmax=6.135e-02; atvarmax(%)=4.000e+01; amflim=(1.150,5.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(4.611e-05); uscref=(4.874e-06)

FieldCalibration MLE SO2

Pandora94s1_FieldCalibration_MLE_SO2.jpeg - 63 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora94s1_FieldCalibration_VCMLE_SO2.jpeg - 58 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=2.905e-04; wrmsmax=5.565e-03; atvarmax(%)=4.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=2; scref=(1.896e-04); uscref=(1.725e-05)

FieldCalibration MLE HCHO

Pandora94s1_FieldCalibration_MLE_HCHO.jpeg - 67 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora94s1_FieldCalibration_VCMLE_HCHO.jpeg - 92 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Unusable
out2 flagged and inverse-u, ous1 also inverse-u

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes
looks good

SO2

Decision for SO2
Yes
looks good, SO2 events

HCHO

Decision for HCHO
Yes
looks good

H2O

Decision for H2O
Yes
looks good