Decision Summary

Upload-ready calibration protocol overview
Instrument P93s1
Session 5
Location LabGSFC
Calibration person Jong-Uk Park
CF file Pandora93s1_CF_v5d20260430.txt
Software version 1.8.85
Calibration vector -
Reference day 20260616
Overall status development
O3Ext: UnusableO3Synt: With CareNO2: YesSO2: With CareHCHO: YesH2O: Not answered
Instrument P93s1
Session 5
Events 538
Figures 17
Action time 6h 24m

Action Timeline

91 action intervals
Jun 25, 2026Jun 26, 2026Jun 27, 2026Jun 28, 2026Jun 29, 202605:55 PM02:57 PMCheckFiltersCheckNoiseWaveAnalyzeDarkAnalyzeBrightWavelengthCalibrationTempDependenceNDFiltersStrayLightSensitivityCalibrationConvolutionCreateReferenceFieldCalibrationAssignCalFileCheckFilters 30m 54s finishedCheckNoiseWave 2m 25s finishedAnalyzeDark 21m 31s finishedAnalyzeDark 1m 23s abortedAnalyzeBright 1h 4m finishedWavelengthCalibration 15m 58s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 7s abortedWavelengthCalibration 7m 33s finishedTempDependence 35s finishedNDFilters 18m 51s finishedNDFilters 1m 0s abortedStrayLight 7m 59s finishedStrayLight 0s finishedStrayLight 0s abortedStrayLight 3m 8s finishedStrayLight 0s finishedStrayLight 0s abortedStrayLight 52s abortedSensitivityCalibration 22m 32s finishedSensitivityCalibration 0s abortedSensitivityCalibration 10s abortedSensitivityCalibration 15s abortedSensitivityCalibration 37m 0s finishedSensitivityCalibration 0s finishedConvolution 35s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedCreateReference 5s abortedCreateReference 15m 29s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 5m 58s finishedCreateReference 0s abortedCreateReference 5m 35s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 5m 8s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 38s finishedCreateReference 0s abortedCreateReference 40s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 44s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 50s finishedCreateReference 0s finishedFieldCalibration 9m 36s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s staleFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 1h 21m finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 18m 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 2m 50s finished

Duration by Step

FieldCalibration1h 49mAnalyzeBright1h 4mSensitivityCalibration59m 57sCreateReference35m 6sCheckFilters30m 54sWavelengthCalibration23m 39sAnalyzeDark22m 53sNDFilters19m 51sStrayLight11m 59sAssignCalFile2m 50sCheckNoiseWave2m 25sConvolution35s

Action Summary

Action Events Runs Aborted Time
AnalyzeBright 14 1 0 1h 4m
AnalyzeDark 12 2 1 22m 53s
AssignCalFile 4 1 0 2m 50s
CheckFilters 4 1 0 30m 54s
CheckNoiseWave 6 1 0 2m 25s
Convolution 16 5 0 35s
CreateReference 50 25 8 35m 6s
FieldCalibration 262 32 0 1h 49m
NDFilters 18 2 1 19m 51s
No action 9 0 0 0s
SensitivityCalibration 77 6 3 59m 57s
StrayLight 37 7 3 11m 59s
TempDependence 7 1 0 35s
WavelengthCalibration 22 6 1 23m 39s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora93s1_WavelengthCalibration_SlitFuncParam.jpeg - 182 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora93s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 171 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora93s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 131 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora93s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 134 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora93s1_CreateReference_AODEST.jpeg - 57 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora93s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 119 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora93s1_CreateReference_DISPFIT(DISP).jpeg - 109 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora93s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 124 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora93s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 147 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora93s1_FieldCalibration_VC_O3.jpeg - 114 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora93s1_FieldCalibration_SCREFINT_O3.jpeg - 84 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 redu=1 settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.000e-04; wrmsmax=5.000e-04; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=30; timeaver(min)=0.00; redu=1; scref=(1.553e-04,1.268e-04,1.354e-04); uscref=(4.633e-06,5.274e-06,8.635e-06)

FieldCalibration MLE NO2

Pandora93s1_FieldCalibration_MLE_NO2.jpeg - 156 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora93s1_FieldCalibration_VCMLE_NO2.jpeg - 130 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.000e-04; wrmsmax=5.000e-04; atvarmax(%)=3.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(2.155e-05); uscref=(4.543e-06)

FieldCalibration MLE SO2

Pandora93s1_FieldCalibration_MLE_SO2.jpeg - 111 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora93s1_FieldCalibration_VCMLE_SO2.jpeg - 94 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO redu=1 settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.000e-03; wrmsmax=1.000e-03; atvarmax(%)=3.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=20; timeaver(min)=0.00; redu=1; scref=(2.457e-04); uscref=(1.793e-05)

FieldCalibration MLE HCHO

Pandora93s1_FieldCalibration_MLE_HCHO.jpeg - 116 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora93s1_FieldCalibration_VCMLE_HCHO.jpeg - 129 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Unusable
Out2 showed strong systematic AMF dependence, presumably given from left shoulder in ISRF. ous1 seemed okay, but from short data availability.

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes
MLE, NO2 results are good

SO2

Decision for SO2
With Care
MLE, Hard to tell the definite quality of SO2 data

HCHO

Decision for HCHO
Yes
MLE, HCHO results are good

H2O

Decision for H2O