Decision Summary

Upload-ready calibration protocol overview
Instrument P90s1
Session 2
Location LabGSFC
Calibration person Apoorva Pandey
CF file Pandora90s1_CF_v2d20241130.txt
Software version 1.8.57
Calibration vector -
Reference day 20250115
Overall status development
O3Ext: YesO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P90s1
Session 2
Events 403
Figures 17
Action time 1h 4m

Action Timeline

92 action intervals
Feb 04, 2025Feb 05, 202506:24 AM05:39 AMCheckFiltersCheckNoiseWaveAnalyzeDarkAnalyzeBrightWavelengthCalibrationTempDependenceNDFiltersSensitivityCalibrationStrayLightConvolutionCreateReferenceFieldCalibrationAssignCalFileCheckFilters 10s finishedCheckNoiseWave 20s finishedAnalyzeDark 4m 31s finishedAnalyzeBright 18m 18s finishedWavelengthCalibration 2m 37s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedTempDependence 51s finishedNDFilters 1m 33s finishedSensitivityCalibration 3s abortedStrayLight 50s finishedStrayLight 0s finishedStrayLight 0s abortedSensitivityCalibration 2m 40s finishedSensitivityCalibration 0s finishedSensitivityCalibration 0s finishedConvolution 1m 22s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedCreateReference 1m 9s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 2m 4s finishedCreateReference 0s abortedCreateReference 1m 37s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 14s abortedCreateReference 1m 5s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 6s finishedCreateReference 0s abortedCreateReference 23s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 27s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 2m 24s finishedCreateReference 0s finishedFieldCalibration 4m 51s abortedFieldCalibration 56s abortedFieldCalibration 5s abortedFieldCalibration 58s finishedFieldCalibration 0s abortedFieldCalibration 44s abortedFieldCalibration 2m 14s finishedFieldCalibration 0s abortedFieldCalibration 16s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 6s abortedFieldCalibration 3m 37s abortedFieldCalibration 3m 22s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 18s abortedFieldCalibration 2m 24s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 22s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 46s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 15s finished

Duration by Step

FieldCalibration21m 0sAnalyzeBright18m 18sCreateReference10m 29sAnalyzeDark4m 31sSensitivityCalibration2m 42sWavelengthCalibration2m 37sNDFilters1m 33sConvolution1m 22sTempDependence51sStrayLight50sCheckNoiseWave20sAssignCalFile15s

Action Summary

Action Events Runs Aborted Time
AnalyzeBright 14 1 0 18m 18s
AnalyzeDark 10 1 0 4m 31s
AssignCalFile 3 1 0 15s
CheckFilters 4 1 0 10s
CheckNoiseWave 6 1 0 20s
Convolution 16 5 0 1m 22s
CreateReference 49 25 8 10m 29s
FieldCalibration 158 43 11 21m 0s
NDFilters 16 1 0 1m 33s
No action 13 0 0 0s
SensitivityCalibration 66 4 1 2m 42s
StrayLight 22 3 1 50s
TempDependence 7 1 0 51s
WavelengthCalibration 19 5 0 2m 37s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora90s1_WavelengthCalibration_SlitFuncParam.jpeg - 129 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora90s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 135 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora90s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 97 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora90s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 90 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora90s1_CreateReference_AODEST.jpeg - 46 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora90s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 95 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora90s1_CreateReference_DISPFIT(DISP).jpeg - 73 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora90s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 103 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora90s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 95 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora90s1_FieldCalibration_VC_O3.jpeg - 81 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora90s1_FieldCalibration_SCREFINT_O3.jpeg - 62 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.198e-04; wrmsmax=2.186e+00; atvarmax(%)=4.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=2; scref=(2.697e-04,2.383e-04,2.721e-04); uscref=(4.285e-05,2.315e-05,2.455e-05)

FieldCalibration MLE NO2

Pandora90s1_FieldCalibration_MLE_NO2.jpeg - 110 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora90s1_FieldCalibration_VCMLE_NO2.jpeg - 73 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(3); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.555e-02; wrmsmax=2.393e-01; atvarmax(%)=4.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=3; timeaver(min)=0.00; redu=0; scref=(1.246e-04); uscref=(9.008e-05)

FieldCalibration MLE SO2

Pandora90s1_FieldCalibration_MLE_SO2.jpeg - 63 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora90s1_FieldCalibration_VCMLE_SO2.jpeg - 52 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(3); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=3.558e-02; wrmsmax=2.767e-01; atvarmax(%)=5.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=30.00; ngroup=5; timeaver(min)=0.00; redu=2; scref=(1.397e-04); uscref=(4.941e-05)

FieldCalibration MLE HCHO

Pandora90s1_FieldCalibration_MLE_HCHO.jpeg - 66 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora90s1_FieldCalibration_VCMLE_HCHO.jpeg - 70 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
AXC-T: clim= 224 K, calculated: 230 K

O3Synt

Decision for O3Synt

NO2

Decision for NO2
E-MLE, NO2 columns and wrms shape looks good, later compared with P2

SO2

Decision for SO2
MLE looks ogood

HCHO

Decision for HCHO
E-MLE, columns look good

H2O

Decision for H2O