Decision Summary
Upload-ready calibration protocol overview
Instrument
P89s1
Session
1
Location
LabGSFC
Calibration person
Apoorva Pandey
CF file
Pandora89s1_CF_v1d20241101.txt
Software version
p1-8
Calibration vector
-
Reference day
20250115
Overall status
development
O3Ext: YesO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Action Timeline
108 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeBright | 14 | 1 | 0 | 1m 38s |
| AnalyzeDark | 14 | 2 | 1 | 4m 2s |
| AssignCalFile | 3 | 1 | 0 | 16s |
| CheckFilters | 4 | 1 | 0 | 6s |
| CheckNoiseWave | 6 | 1 | 0 | 17s |
| Convolution | 16 | 5 | 0 | 1m 3s |
| CreateReference | 107 | 51 | 16 | 34m 51s |
| FieldCalibration | 115 | 32 | 0 | 8m 9s |
| NDFilters | 16 | 1 | 0 | 2m 20s |
| No action | 11 | 0 | 0 | 0s |
| SensitivityCalibration | 70 | 3 | 0 | 5m 56s |
| StrayLight | 20 | 3 | 1 | 29s |
| TempDependence | 7 | 1 | 0 | 27s |
| WavelengthCalibration | 21 | 6 | 1 | 1m 55s |
Session Figures
No session figures found for the loaded instrument/session.No figures loaded
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
AXC-T: clim= 224 K, calculated: 230 K
O3Synt
Decision for O3Synt
NO2
Decision for NO2
E-MLE, NO2 columns and wrms shape looks good, later compared with P2
SO2
Decision for SO2
MLE looks ogood
HCHO
Decision for HCHO
E-MLE, columns look good
H2O