Decision Summary
Upload-ready calibration protocol overview
Instrument
P87s1
Session
1
Location
LabGSFC
Calibration person
Apoorva Pandey
CF file
C:/Blick_old/data/calresults/Pandora87s1/session1/Pandora87s1_CF_session1.txt
Software version
p1-8
Calibration vector
-
Reference day
20240716
Overall status
development
O3Ext: YesO3Synt: With CareNO2: YesSO2: YesHCHO: YesH2O: Not answered
Action Timeline
107 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeBright | 14 | 1 | 0 | 1m 49s |
| AnalyzeDark | 10 | 1 | 0 | 2h 17m |
| CheckFilters | 4 | 1 | 0 | 33s |
| CheckNoiseWave | 6 | 1 | 0 | 1m 39s |
| Convolution | 25 | 8 | 1 | 6m 54s |
| CreateReference | 52 | 26 | 7 | 11m 16s |
| FieldCalibration | 144 | 44 | 10 | 1h 29m |
| NDFilters | 24 | 4 | 3 | 1h 35m |
| No action | 12 | 0 | 0 | 0s |
| SensitivityCalibration | 115 | 6 | 1 | 2m 51s |
| StrayLight | 49 | 9 | 2 | 2m 26s |
| TempDependence | 7 | 1 | 0 | 1m 40s |
| WavelengthCalibration | 19 | 5 | 0 | 2m 9s |
Session Figures
No session figures found for the loaded instrument/session.No figures loaded
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
AXC-T = clim=232.6K, calculated: 238K. In determining SCDref, high stray light contribution (scatter) for AMFs>3
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
E-MLE, NO2 columns and wrms shape looks good, later compared with P2
SO2
Decision for SO2
Yes
Columns look okay, but instrument temperature problems on clear days resulted in a smaller dataset than typical for clearance
HCHO
Decision for HCHO
Yes
Columns look okay, but instrument temperature problems on clear days resulted in a smaller dataset than typical for clearance
H2O
Decision for H2O