Decision Summary
Upload-ready calibration protocol overview
Instrument
P85s1
Session
3
Location
LabGSFC
Calibration person
Apoorva Pandey
CF file
C:/Blick_old/data/calresults/Pandora85s1/session3/Pandora85s1_CF_session3.txt
Software version
p1-8
Calibration vector
-
Reference day
20250409
Overall status
development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Action Timeline
83 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeBright | 14 | 1 | 0 | 24m 28s |
| AnalyzeDark | 14 | 2 | 1 | 41m 54s |
| CheckFilters | 16 | 5 | 4 | 33s |
| CheckNoiseWave | 6 | 1 | 0 | 9s |
| Convolution | 16 | 5 | 0 | 47s |
| CreateReference | 52 | 26 | 7 | 3h 0m |
| FieldCalibration | 93 | 27 | 2 | 1h 44m |
| NDFilters | 16 | 1 | 0 | 2m 16s |
| No action | 14 | 0 | 0 | 0s |
| SensitivityCalibration | 64 | 3 | 0 | 2m 38s |
| StrayLight | 25 | 4 | 2 | 48s |
| TempDependence | 7 | 1 | 0 | 25s |
| WavelengthCalibration | 22 | 6 | 1 | 3m 15s |
Session Figures
15 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora85s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 129 KBCreate reference SELSPECS
Pandora85s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 96 KBCreate reference SLERR (STRAY)
Pandora85s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 105 KBCreate reference AODEST
Pandora85s1_CreateReference_AODEST.jpeg - 42 KBCreate reference SENSCORRFIN (SENS)
Pandora85s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 114 KBCreate reference DISPFIT (DISP)
Pandora85s1_CreateReference_DISPFIT(DISP).jpeg - 123 KBCreate reference WLCHANGE (DISP)
Pandora85s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 133 KBCreate reference SHIFT (CLOS)
Pandora85s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 122 KB
field calibration
O3
field calibration
NO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
AXC looks good
O3Synt
Decision for O3Synt
NO2
Decision for NO2
E-MLE, NO2 columns and wrms shape looks good, later compared with P2
SO2
Decision for SO2
HCHO
Decision for HCHO
E-MLE looks good
H2O
Decision for H2O