Decision Summary
Upload-ready calibration protocol overview
Instrument
P84s1
Session
11
Location
LabGSFC
Calibration person
Jong-Uk Park
CF file
Pandora84s1_CF_v11d20260319.txt
Software version
1.8.85
Calibration vector
-
Reference day
20260324
Overall status
development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: Not answeredHCHO: With CareH2O: Not answered
Action Timeline
112 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeBright | 36 | 5 | 3 | 39m 54s |
| AnalyzeDark | 12 | 2 | 1 | 1m 11s |
| AssignCalFile | 4 | 1 | 0 | 6m 47s |
| CheckFilters | 4 | 1 | 0 | 1m 40s |
| CheckNoiseWave | 11 | 2 | 0 | 1m 8s |
| Convolution | 16 | 5 | 0 | 1m 0s |
| CreateReference | 64 | 31 | 11 | 31m 51s |
| FieldCalibration | 122 | 35 | 5 | 13m 42s |
| NDFilters | 52 | 5 | 2 | 13m 25s |
| No action | 27 | 0 | 0 | 0s |
| SensitivityCalibration | 104 | 8 | 4 | 10m 13s |
| StrayLight | 41 | 8 | 4 | 17m 14s |
| TempDependence | 7 | 1 | 0 | 1m 4s |
| WavelengthCalibration | 21 | 5 | 0 | 25m 59s |
Session Figures
11 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora84s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 158 KBCreate reference SELSPECS
Pandora84s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 127 KBCreate reference SLERR (STRAY)
Pandora84s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 135 KBCreate reference AODEST
Pandora84s1_CreateReference_AODEST.jpeg - 57 KBCreate reference SENSCORRFIN (SENS)
Pandora84s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 118 KBCreate reference DISPFIT (DISP)
Pandora84s1_CreateReference_DISPFIT(DISP).jpeg - 112 KBCreate reference WLCHANGE (DISP)
Pandora84s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 198 KBCreate reference SHIFT (CLOS)
Pandora84s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 134 KB
field calibration
O3
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
AXC-T = clim 227.06K, calculated 229.03 K. out2 wrms usually in between DQ1 and DQ2 thresholds. ous1 showing abnormal wrms behavior - higher around noon.
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
Ref method. matched well with reference
SO2
Decision for SO2
HCHO
Decision for HCHO
With Care
Notable difference between total column and tropospheric column (e.g., offset)
H2O