Decision Summary
Upload-ready calibration protocol overview
Instrument
P82s1
Session
1
Location
LabGSFC
Calibration person
Apoorva Pandey
CF file
C:/Blick/data/calresults/Pandora82s1/session1/Pandora82s1_CF_session1.txt
Software version
p1-8
Calibration vector
-
Reference day
20240423
Overall status
development
O3Ext: YesO3Synt: YesNO2: YesSO2: YesHCHO: YesH2O: Not answered
Action Timeline
119 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeBright | 14 | 1 | 0 | 8m 7s |
| AnalyzeDark | 13 | 2 | 1 | 2h 2m |
| CheckFilters | 18 | 6 | 5 | 46m 1s |
| CheckNoiseWave | 9 | 2 | 1 | 45s |
| Convolution | 16 | 5 | 0 | 9m 57s |
| CreateReference | 81 | 39 | 11 | 31m 57s |
| FieldCalibration | 184 | 33 | 11 | 34m 43s |
| NDFilters | 18 | 1 | 0 | 4m 2s |
| No action | 20 | 0 | 0 | 0s |
| SensitivityCalibration | 158 | 7 | 2 | 15m 44s |
| StrayLight | 51 | 11 | 5 | 8m 2s |
| TempDependence | 7 | 1 | 0 | 34s |
| WavelengthCalibration | 33 | 9 | 4 | 8m 32s |
Session Figures
11 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora82s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 126 KBCreate reference SELSPECS
Pandora82s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 100 KBCreate reference SLERR (STRAY)
Pandora82s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 100 KBCreate reference AODEST
Pandora82s1_CreateReference_AODEST.jpeg - 42 KBCreate reference SENSCORRFIN (SENS)
Pandora82s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 90 KBCreate reference DISPFIT (DISP)
Pandora82s1_CreateReference_DISPFIT(DISP).jpeg - 90 KBCreate reference WLCHANGE (DISP)
Pandora82s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 97 KBCreate reference SHIFT (CLOS)
Pandora82s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 115 KB
field calibration
O3
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
O3T within 10K of climatology
O3Synt
Decision for O3Synt
Yes
NO2
Decision for NO2
Yes
NO2 columns look good, wrms is a little noisy but very low, and comes down close to expected wrms
SO2
Decision for SO2
Yes
matched well with reference
HCHO
Decision for HCHO
Yes
matched well with trop. columns
H2O
Decision for H2O