Decision Summary
Upload-ready calibration protocol overview
Instrument
P78s1
Session
3
Location
Banting
Calibration person
Stefanie Morhenn
CF file
Pandora78s1_CF_v3d20220622.txt
Software version
1.8.88
Calibration vector
CalV1
Reference day
20240322
Overall status
development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: YesHCHO: YesH2O: Unusable
Action Timeline
109 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 3 | 1 | 0 | 1h 3m |
| Convolution | 20 | 7 | 2 | 2m 4s |
| CreateReference | 93 | 45 | 16 | 2h 43m |
| FieldCalibration | 411 | 45 | 4 | 1h 3m |
| NDFilters | 50 | 1 | 0 | 2m 2s |
| No action | 9 | 0 | 0 | 0s |
| SensitivityCalibration | 78 | 7 | 3 | 1m 32s |
| WavelengthCalibration | 5 | 2 | 1 | 16m 49s |
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora78s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 148 KBCreate reference SELSPECS
Pandora78s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 123 KBCreate reference SLERR (STRAY)
Pandora78s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 224 KBCreate reference AODEST
Pandora78s1_CreateReference_AODEST.jpeg - 65 KBCreate reference SENSCORRFIN (SENS)
Pandora78s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 126 KBCreate reference DISPFIT (DISP)
Pandora78s1_CreateReference_DISPFIT(DISP).jpeg - 108 KBCreate reference WLCHANGE (DISP)
Pandora78s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 114 KBCreate reference SHIFT (CLOS)
Pandora78s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 111 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
slight AMF dependency
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
looks good, but sky data flagged out
SO2
Decision for SO2
Yes
looks good
HCHO
Decision for HCHO
Yes
looks good, but sky data flagged out
H2O
Decision for H2O
Unusable
flagged