Decision Summary
Upload-ready calibration protocol overview
Instrument
P75s1
Session
2
Location
LabGSFC
Calibration person
Bryan Place
CF file
Pandora75s1_CF_v2d20230525.txt
Software version
1.8.56
Calibration vector
-
Reference day
20230902
Overall status
development
O3Ext: UnusableO3Synt: YesNO2: YesSO2: YesHCHO: YesH2O: Not answered
Action Timeline
104 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeBright | 14 | 1 | 0 | 6m 32s |
| AnalyzeDark | 18 | 3 | 1 | 4m 24s |
| AssignCalFile | 3 | 1 | 0 | 4m 34s |
| CheckFilters | 6 | 2 | 1 | 31s |
| CheckNoiseWave | 8 | 2 | 1 | 41s |
| Convolution | 16 | 5 | 0 | 1m 19s |
| CreateReference | 61 | 31 | 13 | 15m 13s |
| FieldCalibration | 131 | 38 | 4 | 1h 33m |
| NDFilters | 18 | 2 | 1 | 2m 1s |
| No action | 16 | 0 | 0 | 0s |
| SensitivityCalibration | 66 | 4 | 1 | 4m 50s |
| StrayLight | 32 | 6 | 4 | 3m 8s |
| TempDependence | 9 | 2 | 1 | 1m 37s |
| WavelengthCalibration | 21 | 6 | 1 | 3m 5s |
Session Figures
11 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora75s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 125 KBCreate reference SELSPECS
Pandora75s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 90 KBCreate reference SLERR (STRAY)
Pandora75s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 90 KBCreate reference AODEST
Pandora75s1_CreateReference_AODEST.jpeg - 37 KBCreate reference SENSCORRFIN (SENS)
Pandora75s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 69 KBCreate reference DISPFIT (DISP)
Pandora75s1_CreateReference_DISPFIT(DISP).jpeg - 64 KBCreate reference WLCHANGE (DISP)
Pandora75s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 67 KBCreate reference SHIFT (CLOS)
Pandora75s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 87 KB
field calibration
O3
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Unusable
Unit suffers from high straylight. Ous1 retrievals and wrms shape look good though
O3Synt
Decision for O3Synt
Yes
NO2
Decision for NO2
Yes
Calibrated with Ref method (P2). NO2 columns follow reference well and wrms shape fine
SO2
Decision for SO2
Yes
Calibrated with Ref method (P2). SO2 columns follow reference well and wrms shape fine
HCHO
Decision for HCHO
Yes
Calibrated with Ref method (P2). HCHO columns follow reference excellently and wrms shape fine
H2O
Decision for H2O