Decision Summary

Upload-ready calibration protocol overview
Instrument P74s1
Session 7
Location -
Calibration person Ragi Rajagopalan
CF file Pandora74s1_CF_v7d20231113.txt
Software version 1.8.66
Calibration vector -
Reference day -
Overall status development
O3Ext: YesO3Synt: YesNO2: YesSO2: YesHCHO: With CareH2O: Not answered
Instrument P74s1
Session 7
Events 12
Figures 17
Action time 35s

Action Timeline

1 action intervals
Aug 22, 202410:02 AM10:18 AMAssignCalFileAssignCalFile 35s finished

Duration by Step

AssignCalFile35s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 35s
No action 9 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora74s1_WavelengthCalibration_SlitFuncParam.jpeg - 163 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora74s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 158 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora74s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 133 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora74s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 117 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora74s1_CreateReference_AODEST.jpeg - 65 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora74s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 124 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora74s1_CreateReference_DISPFIT(DISP).jpeg - 98 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora74s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 133 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora74s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 113 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora74s1_FieldCalibration_VC_O3.jpeg - 96 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora74s1_FieldCalibration_SCREFINT_O3.jpeg - 76 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora74s1_FieldCalibration_MLE_NO2.jpeg - 210 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora74s1_FieldCalibration_VCMLE_NO2.jpeg - 123 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora74s1_FieldCalibration_MLE_SO2.jpeg - 109 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora74s1_FieldCalibration_VCMLE_SO2.jpeg - 101 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora74s1_FieldCalibration_MLE_HCHO.jpeg - 164 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora74s1_FieldCalibration_VCMLE_HCHO.jpeg - 141 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
Some inverted U-shape persistent in O3Ext nd Temperature but looks okay otherwise, probably one of the instruments which needs Lorentz as slit function

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
Seems good with New VP

SO2

Decision for SO2
Yes
looks okay, but some ´days have weird V-shape

HCHO

Decision for HCHO
With Care
For some days, the tropspheric values are higher. Was an issue with HCHO calibration, the tropspheric had inverted U shape. Used MLE with perc 50 and adjusted scref to get better total columns.

H2O

Decision for H2O
Not Calibrated