Decision Summary
Upload-ready calibration protocol overview
Instrument
P74s1
Session
7
Location
-
Calibration person
Ragi Rajagopalan
CF file
Pandora74s1_CF_v7d20231113.txt
Software version
1.8.66
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: YesO3Synt: YesNO2: YesSO2: YesHCHO: With CareH2O: Not answered
Action Timeline
1 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 3 | 1 | 0 | 35s |
| No action | 9 | 0 | 0 | 0s |
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora74s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 158 KBCreate reference SELSPECS
Pandora74s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 133 KBCreate reference SLERR (STRAY)
Pandora74s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 117 KBCreate reference AODEST
Pandora74s1_CreateReference_AODEST.jpeg - 65 KBCreate reference SENSCORRFIN (SENS)
Pandora74s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 124 KBCreate reference DISPFIT (DISP)
Pandora74s1_CreateReference_DISPFIT(DISP).jpeg - 98 KBCreate reference WLCHANGE (DISP)
Pandora74s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 133 KBCreate reference SHIFT (CLOS)
Pandora74s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 113 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
Some inverted U-shape persistent in O3Ext nd Temperature but looks okay otherwise, probably one of the instruments which needs Lorentz as slit function
O3Synt
Decision for O3Synt
Yes
NO2
Decision for NO2
Yes
Seems good with New VP
SO2
Decision for SO2
Yes
looks okay, but some ´days have weird V-shape
HCHO
Decision for HCHO
With Care
For some days, the tropspheric values are higher. Was an issue with HCHO calibration, the tropspheric had inverted U shape. Used MLE with perc 50 and adjusted scref to get better total columns.
H2O
Decision for H2O
Not Calibrated