Decision Summary

Upload-ready calibration protocol overview
Instrument P70s1
Session 30
Location -
Calibration person -
CF file -
Software version 1.8.50
Calibration vector -
Reference day -
Overall status development
O3Ext: YesO3Synt: YesNO2: YesSO2: YesHCHO: With CareH2O: Yes
Instrument P70s1
Session 30
Events 0
Figures 17
Action time 0s

Action Timeline

No start/finish intervals found.
No timeline intervals found

Duration by Step

No completed durations found

Action Summary

Action Events Runs Aborted Time
No actions loaded

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora70s1_WavelengthCalibration_SlitFuncParam.jpeg - 123 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora70s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 232 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora70s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 182 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora70s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 185 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora70s1_CreateReference_AODEST.jpeg - 89 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora70s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 167 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora70s1_CreateReference_DISPFIT(DISP).jpeg - 152 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora70s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 201 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora70s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 199 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora70s1_FieldCalibration_VC_O3.jpeg - 157 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora70s1_FieldCalibration_SCREFINT_O3.jpeg - 116 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora70s1_FieldCalibration_MLE_NO2.jpeg - 290 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora70s1_FieldCalibration_VCMLE_NO2.jpeg - 182 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora70s1_FieldCalibration_MLE_SO2.jpeg - 153 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora70s1_FieldCalibration_VCMLE_SO2.jpeg - 133 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora70s1_FieldCalibration_MLE_HCHO.jpeg - 146 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora70s1_FieldCalibration_VCMLE_HCHO.jpeg - 172 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
ok

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
EMLE used

SO2

Decision for SO2
Yes
ok

HCHO

Decision for HCHO
With Care
seems to be ok until end of 2022 then wiggles increase in magnitude

H2O

Decision for H2O
Yes
ok