Decision Summary

Upload-ready calibration protocol overview
Instrument P63s1
Session 9
Location -
Calibration person SMorhenn
CF file -
Software version 1.8.75
Calibration vector -
Reference day -
Overall status development
O3Ext: UnusableO3Synt: UnusableNO2: YesSO2: With CareHCHO: With CareH2O: Yes
Instrument P63s1
Session 9
Events 0
Figures 17
Action time 0s

Action Timeline

No start/finish intervals found.
No timeline intervals found

Duration by Step

No completed durations found

Action Summary

Action Events Runs Aborted Time
No actions loaded

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora63s1_WavelengthCalibration_SlitFuncParam.jpeg - 78 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora63s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 164 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora63s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 123 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora63s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 119 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora63s1_CreateReference_AODEST.jpeg - 69 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora63s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 174 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora63s1_CreateReference_DISPFIT(DISP).jpeg - 153 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora63s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 162 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora63s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 165 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora63s1_FieldCalibration_VC_O3.jpeg - 150 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora63s1_FieldCalibration_SCREFINT_O3.jpeg - 112 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora63s1_FieldCalibration_MLE_NO2.jpeg - 172 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora63s1_FieldCalibration_VCMLE_NO2.jpeg - 104 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora63s1_FieldCalibration_MLE_SO2.jpeg - 114 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora63s1_FieldCalibration_VCMLE_SO2.jpeg - 101 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora63s1_FieldCalibration_MLE_HCHO.jpeg - 98 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora63s1_FieldCalibration_VCMLE_HCHO.jpeg - 111 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Unusable
O3T taken from climatlogy in AXC: too low (u-shape)

O3Synt

Decision for O3Synt
Unusable

NO2

Decision for NO2
Yes
looks good. E-MLE used although many sky data points flagged out

SO2

Decision for SO2
With Care
inverse-u

HCHO

Decision for HCHO
With Care
too high in winter, too low in summer

H2O

Decision for H2O
Yes
ok