Decision Summary

Upload-ready calibration protocol overview
Instrument P56s1
Session 8
Location MaunaLoaHI
Calibration person Markus Kilian
CF file Pandora56s1_CF_v8d20200803.txt
Software version p1-8
Calibration vector -
Reference day 20200829
Overall status development
O3Ext: YesO3Synt: With CareNO2: YesSO2: YesHCHO: YesH2O: Yes
Instrument P56s1
Session 8
Events 308
Figures 0
Action time 56m 12s

Action Timeline

80 action intervals
Jan 31, 202509:12 AM11:38 AMFieldCalibrationWavelengthCalibrationCheckFiltersCreateReferenceAssignCalFileFieldCalibration 4s abortedWavelengthCalibration 12s finishedCheckFilters 14s abortedCreateReference 18s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 10s finishedCreateReference 0s abortedCreateReference 13s finishedCreateReference 0s abortedCreateReference 57s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 32s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 16s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 22s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 19s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 15s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 19s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 31s finishedCreateReference 0s finishedFieldCalibration 1m 33s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 15s abortedFieldCalibration 55s finishedFieldCalibration 0s abortedFieldCalibration 25s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 21m 58s finishedFieldCalibration 0s abortedFieldCalibration 1m 58s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 8m 59s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 11m 0s finishedFieldCalibration 0s abortedFieldCalibration 8s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 3m 17s finished

Duration by Step

FieldCalibration47m 15sCreateReference5m 14sAssignCalFile3m 17sCheckFilters14sWavelengthCalibration12s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 3m 17s
CheckFilters 2 1 1 14s
CreateReference 74 36 10 5m 14s
FieldCalibration 213 41 6 47m 15s
No action 13 0 0 0s
WavelengthCalibration 3 1 0 12s

Session Figures

No session figures found for the loaded instrument/session.
No figures loaded

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
looks ok, O3Synt with invers u-shape

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes
E-MLE looks good

SO2

Decision for SO2
Yes
wiggly, but okay

HCHO

Decision for HCHO
Yes
okay, when alignment was fine

H2O

Decision for H2O
Yes
quality ok, but high wl_shift, reason could not be found.