Decision Summary

Upload-ready calibration protocol overview
Instrument P54s1
Session 33
Location LabNIER
Calibration person Hanna Rohringer
CF file C:/Blick-dev18/data/calresults/Pandora54s1/session33/Pandora54s1_CF_session33.txt
Software version 1.8.89
Calibration vector -
Reference day -
Overall status development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: YesHCHO: YesH2O: Yes
Instrument P54s1
Session 33
Events 102
Figures 1
Action time 2h 52m

Action Timeline

21 action intervals
May 11, 202609:18 AM02:06 PMSensitivityCalibrationWavelengthCalibrationNDFiltersSensitivityCalibration 15s finishedSensitivityCalibration 0s finishedSensitivityCalibration 1s abortedWavelengthCalibration 3m 52s abortedWavelengthCalibration 6s abortedWavelengthCalibration 21s abortedWavelengthCalibration 8m 10s abortedWavelengthCalibration 1m 44s abortedWavelengthCalibration 7s abortedWavelengthCalibration 2m 45s abortedWavelengthCalibration 37s abortedWavelengthCalibration 1h 22m abortedWavelengthCalibration 9s abortedWavelengthCalibration 53m 52s abortedWavelengthCalibration 4m 1s abortedWavelengthCalibration 1m 14s abortedWavelengthCalibration 32s finishedWavelengthCalibration 9m 16s abortedWavelengthCalibration 32s abortedNDFilters 3m 2s abortedWavelengthCalibration 1s aborted

Duration by Step

WavelengthCalibration2h 49mNDFilters3m 2sSensitivityCalibration16s

Action Summary

Action Events Runs Aborted Time
AnalyzeDark 1 0 0 0s
Convolution 1 0 0 0s
NDFilters 2 1 1 3m 2s
No action 12 0 0 0s
SensitivityCalibration 38 3 1 16s
WavelengthCalibration 48 17 16 2h 49m

Session Figures

1 figure from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora54s1_WavelengthCalibration_SlitFuncParam.jpeg - 124 KB
Slit function parameters

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
With Care
O3Synt looks good. O3Ext almost everything flagged out

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes
looks good, E-MLE used

SO2

Decision for SO2
Yes
looks good, few days with inverse u-shape

HCHO

Decision for HCHO
Yes
looks good, E-MLE used

H2O

Decision for H2O
Yes
looks good