Decision Summary
Upload-ready calibration protocol overview
Instrument
P54s1
Session
33
Location
LabNIER
Calibration person
Hanna Rohringer
CF file
C:/Blick-dev18/data/calresults/Pandora54s1/session33/Pandora54s1_CF_session33.txt
Software version
1.8.89
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: YesHCHO: YesH2O: Yes
Action Timeline
21 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeDark | 1 | 0 | 0 | 0s |
| Convolution | 1 | 0 | 0 | 0s |
| NDFilters | 2 | 1 | 1 | 3m 2s |
| No action | 12 | 0 | 0 | 0s |
| SensitivityCalibration | 38 | 3 | 1 | 16s |
| WavelengthCalibration | 48 | 17 | 16 | 2h 49m |
Session Figures
1 figure from Google Drive/figuresAutomatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
O3Synt looks good. O3Ext almost everything flagged out
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
looks good, E-MLE used
SO2
Decision for SO2
Yes
looks good, few days with inverse u-shape
HCHO
Decision for HCHO
Yes
looks good, E-MLE used
H2O
Decision for H2O
Yes
looks good