Decision Summary
Upload-ready calibration protocol overview
Instrument
P34s1
Session
3
Location
LabGSFC
Calibration person
Fernando Santos
CF file
Pandora34s1_CF_v3d20230605.txt
Software version
1.8.57
Calibration vector
-
Reference day
20230727
Overall status
development
O3Ext: YesO3Synt: YesNO2: YesSO2: YesHCHO: With CareH2O: Yes
Action Timeline
122 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 7 | 2 | 0 | 7m 20s |
| CheckDispersion | 21 | 6 | 0 | 5m 39s |
| CheckFilters | 2 | 1 | 1 | 4s |
| Convolution | 16 | 5 | 0 | 39s |
| CreateReference | 165 | 43 | 13 | 20m 24s |
| FieldCalibration | 210 | 46 | 10 | 3h 37m |
| NDFilters | 298 | 5 | 2 | 1h 22m |
| No action | 31 | 0 | 0 | 0s |
| SensitivityCalibration | 168 | 6 | 0 | 8m 10s |
| TempDependence | 4 | 2 | 2 | 13s |
| WavelengthCalibration | 8 | 3 | 1 | 22m 47s |
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora34s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 231 KBCreate reference SELSPECS
Pandora34s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 185 KBCreate reference SLERR (STRAY)
Pandora34s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 175 KBCreate reference AODEST
Pandora34s1_CreateReference_AODEST.jpeg - 84 KBCreate reference SENSCORRFIN (SENS)
Pandora34s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 170 KBCreate reference DISPFIT (DISP)
Pandora34s1_CreateReference_DISPFIT(DISP).jpeg - 137 KBCreate reference WLCHANGE (DISP)
Pandora34s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 281 KBCreate reference SHIFT (CLOS)
Pandora34s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 136 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
no extrapolation or interpolation needed
O3Synt
Decision for O3Synt
Yes
NO2
Decision for NO2
Yes
EMLE, close to first gues
SO2
Decision for SO2
Yes
very low values so has a weird shape
HCHO
Decision for HCHO
With Care
some days are less than trop
H2O
Decision for H2O
Yes