Decision Summary

Upload-ready calibration protocol overview
Instrument P307s1
Session 1
Location LabGSFC
Calibration person Bryan Place
CF file Pandora307s1_CF_v1d20251230.txt
Software version 1.8.93
Calibration vector -
Reference day 20260112
Overall status development
O3Ext: YesO3Synt: YesNO2: YesSO2: With CareHCHO: YesH2O: Not answered
Instrument P307s1
Session 1
Events 420
Figures 17
Action time 5h 20m

Action Timeline

86 action intervals
Feb 04, 2026Feb 05, 2026Feb 06, 2026May 05, 202607:59 PMCheckFiltersCheckNoiseWaveAnalyzeDarkAnalyzeBrightWavelengthCalibrationTempDependenceNDFiltersStrayLightSensitivityCalibrationConvolutionCreateReferenceFieldCalibrationAssignCalFileCheckFilters 24s finishedCheckNoiseWave 17s finishedAnalyzeDark 7m 28s finishedAnalyzeBright 3m 29s finishedWavelengthCalibration 1m 51s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedTempDependence 37s finishedNDFilters 3m 42s finishedStrayLight 1m 15s finishedStrayLight 0s finishedStrayLight 0s abortedSensitivityCalibration 2m 37s finishedSensitivityCalibration 0s finishedSensitivityCalibration 0s finishedConvolution 1m 19s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedCreateReference 28s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 42s finishedCreateReference 0s abortedCreateReference 3m 29s abortedCreateReference 56s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 45m 24s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 23s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 4s abortedCreateReference 20s abortedCreateReference 37s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 32s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 44s finishedCreateReference 0s finishedFieldCalibration 7m 2s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 1h 9m finishedFieldCalibration 0s abortedFieldCalibration 38s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 2h 37m finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 7m 55s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedCreateReference 9s abortedAssignCalFile 13s finished

Duration by Step

FieldCalibration4h 3mCreateReference53m 50sAnalyzeDark7m 28sNDFilters3m 42sAnalyzeBright3m 29sSensitivityCalibration2m 37sWavelengthCalibration1m 51sConvolution1m 19sStrayLight1m 15sTempDependence37sCheckFilters24sCheckNoiseWave17s

Action Summary

Action Events Runs Aborted Time
AnalyzeBright 14 1 0 3m 29s
AnalyzeDark 10 1 0 7m 28s
AssignCalFile 4 1 0 13s
CheckFilters 4 1 0 24s
CheckNoiseWave 6 1 0 17s
Convolution 16 5 0 1m 19s
CreateReference 60 30 11 53m 50s
FieldCalibration 161 33 1 4h 3m
NDFilters 16 1 0 3m 42s
No action 17 0 0 0s
SensitivityCalibration 64 3 0 2m 37s
StrayLight 22 3 1 1m 15s
TempDependence 7 1 0 37s
WavelengthCalibration 19 5 0 1m 51s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora307s1_WavelengthCalibration_SlitFuncParam.jpeg - 119 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora307s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 111 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora307s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 81 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora307s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 88 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora307s1_CreateReference_AODEST.jpeg - 37 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora307s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 83 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora307s1_CreateReference_DISPFIT(DISP).jpeg - 78 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora307s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 128 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora307s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 98 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora307s1_FieldCalibration_VC_O3.jpeg - 66 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora307s1_FieldCalibration_SCREFINT_O3.jpeg - 51 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.491e-04; wrmsmax=3.184e+00; atvarmax(%)=5.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=2; scref=(3.088e-04,3.022e-04,3.177e-04); uscref=(1.886e-05,3.696e-05,2.832e-05)

FieldCalibration MLE NO2

Pandora307s1_FieldCalibration_MLE_NO2.jpeg - 82 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora307s1_FieldCalibration_VCMLE_NO2.jpeg - 54 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.452e-02; wrmsmax=2.665e-01; atvarmax(%)=2.000e+01; amflim=(1.000,2.500); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(4.664e-05); uscref=(4.669e-05)

FieldCalibration MLE SO2

Pandora307s1_FieldCalibration_MLE_SO2.jpeg - 56 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora307s1_FieldCalibration_VCMLE_SO2.jpeg - 46 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=2.342e-02; wrmsmax=2.582e-01; atvarmax(%)=7.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=20.00; ngroup=10; timeaver(min)=0.00; redu=2; scref=(1.318e-04); uscref=(6.169e-05)

FieldCalibration MLE HCHO

Pandora307s1_FieldCalibration_MLE_HCHO.jpeg - 55 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora307s1_FieldCalibration_VCMLE_HCHO.jpeg - 58 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
AXC-T = clim:223 K, calculated: 224.6K. All O3 products looked good, wrms below limits and followed same pattern as P2

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
E-MLE used; NO2 columns and wrms look good.

SO2

Decision for SO2
With Care
MLE used; SO2 columns look OK, slight AMF shape but scatter around 0

HCHO

Decision for HCHO
Yes
E-MLE used; HCHO looks good

H2O

Decision for H2O