Decision Summary

Upload-ready calibration protocol overview
Instrument P29s1
Session 12
Location FairbanksAK
Calibration person Stefanie Morhenn
CF file Pandora29s1_CF_v12d20170504.txt
Software version 1.8.75
Calibration vector CalV3
Reference day -
Overall status development
O3Ext: YesO3Synt: UnusableNO2: YesSO2: YesHCHO: YesH2O: Yes
Instrument P29s1
Session 12
Events 83
Figures 17
Action time 33m 44s

Action Timeline

12 action intervals
Apr 14, 202509:42 AM03:38 PMFieldCalibrationAssignCalFileFieldCalibration 13m 8s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 20m 9s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 27s finished

Duration by Step

FieldCalibration33m 17sAssignCalFile27s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 27s
FieldCalibration 71 11 0 33m 17s
No action 9 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora29s1_WavelengthCalibration_SlitFuncParam.jpeg - 249 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora29s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 164 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora29s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 132 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora29s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 115 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora29s1_CreateReference_AODEST.jpeg - 65 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora29s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 114 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora29s1_CreateReference_DISPFIT(DISP).jpeg - 109 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora29s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 123 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora29s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 134 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora29s1_FieldCalibration_VC_O3.jpeg - 100 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora29s1_FieldCalibration_SCREFINT_O3.jpeg - 79 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora29s1_FieldCalibration_MLE_NO2.jpeg - 170 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora29s1_FieldCalibration_VCMLE_NO2.jpeg - 124 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora29s1_FieldCalibration_MLE_SO2.jpeg - 131 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora29s1_FieldCalibration_VCMLE_SO2.jpeg - 113 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO MLE settings applied fw1out=(); fw2out=(); dqf2in=(103); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=4.303e-02; wrmsmax=9.000e-04; atvarmax(%)=6.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=10.00; ngroup=62; timeaver(min)=0.00; redu=0; scref=(1.410e-04); uscref=(8.235e-06)

FieldCalibration MLE HCHO

Pandora29s1_FieldCalibration_MLE_HCHO.jpeg - 215 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora29s1_FieldCalibration_VCMLE_HCHO.jpeg - 207 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
O3T taken from climatology (cto0 was 20 K above clim)

O3Synt

Decision for O3Synt
Unusable

NO2

Decision for NO2
Yes
looks good

SO2

Decision for SO2
Yes
looks good

HCHO

Decision for HCHO
Yes
looks good

H2O

Decision for H2O
Yes
looks good