Decision Summary
Upload-ready calibration protocol overview
Instrument
P297s1
Session
2
Location
LabGSFC
Calibration person
Bryan Place
CF file
Pandora297s1_CF_v2d20250814.txt
Software version
p1-8
Calibration vector
-
Reference day
20250822
Overall status
development
O3Ext: YesO3Synt: YesNO2: YesSO2: YesHCHO: YesH2O: Not answered
Action Timeline
91 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeBright | 20 | 3 | 2 | 6m 13s |
| AnalyzeDark | 14 | 3 | 2 | 2m 53s |
| AssignCalFile | 4 | 1 | 0 | 30s |
| CheckFilters | 6 | 2 | 1 | 21s |
| CheckNoiseWave | 6 | 1 | 0 | 37s |
| Convolution | 16 | 5 | 0 | 1m 0s |
| CreateReference | 43 | 22 | 6 | 8m 4s |
| FieldCalibration | 128 | 35 | 2 | 1h 10m |
| NDFilters | 36 | 2 | 1 | 2m 41s |
| No action | 19 | 0 | 0 | 0s |
| SensitivityCalibration | 78 | 4 | 1 | 3m 44s |
| StrayLight | 24 | 3 | 1 | 34s |
| TempDependence | 9 | 2 | 1 | 33s |
| WavelengthCalibration | 23 | 7 | 2 | 3m 22s |
Session Figures
No session figures found for the loaded instrument/session.No figures loaded
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
AXC-T clim: 231.3K calculated: 231.7K. Out2 and Ous1 shape looked good and wrms below limits. Matched REF columns well.
O3Synt
Decision for O3Synt
Yes
NO2
Decision for NO2
Yes
Calibrated with REF method. Columns matched P2 well and no obvious shape in wrms
SO2
Decision for SO2
Yes
Calibrated with REF method and columns tracked very well with P2
HCHO
Decision for HCHO
Yes
Calibrated with REF method and columns tracked very well with P2
H2O
Decision for H2O