Decision Summary

Upload-ready calibration protocol overview
Instrument P296s1
Session 4
Location Aosta-SC
Calibration person Ragi Rajagopalan
CF file Pandora296s1_CF_v4d20251001.txt
Software version 1.8.94
Calibration vector calV1
Reference day 20260323
Overall status development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: YesHCHO: YesH2O: Yes
Instrument P296s1
Session 4
Events 243
Figures 17
Action time 1h 51m

Action Timeline

76 action intervals
Mar 25, 202611:04 AM06:24 PMCreateReferenceFieldCalibrationAssignCalFileCreateReference 17s finishedCreateReference 0s abortedCreateReference 42s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 2m 45s finishedCreateReference 0s abortedCreateReference 1m 24s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 7s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 33s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 20s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 45s finishedCreateReference 0s abortedCreateReference 18s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 24s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 25s finishedCreateReference 0s finishedCreateReference 3m 12s finishedFieldCalibration 12s abortedFieldCalibration 2s abortedFieldCalibration 2s abortedCreateReference 6s abortedFieldCalibration 50s abortedFieldCalibration 3m 14s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 1h 9m finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 40s abortedFieldCalibration 7m 51s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 1m 42s finishedFieldCalibration 0s finishedFieldCalibration 14m 38s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 52s finished

Duration by Step

FieldCalibration1h 38mCreateReference12m 19sAssignCalFile52s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 52s
CreateReference 74 36 11 12m 19s
FieldCalibration 153 39 5 1h 38m
No action 13 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora296s1_WavelengthCalibration_SlitFuncParam.jpeg - 166 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora296s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 105 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora296s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 77 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora296s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 126 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora296s1_CreateReference_AODEST.jpeg - 39 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora296s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 73 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora296s1_CreateReference_DISPFIT(DISP).jpeg - 69 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora296s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 67 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora296s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 24 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora296s1_FieldCalibration_VC_O3.jpeg - 60 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora296s1_FieldCalibration_SCREFINT_O3.jpeg - 46 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=6.614e-05; wrmsmax=2.152e+00; atvarmax(%)=2.000e+01; amflim=(1.000,3.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=14; timeaver(min)=0.00; redu=2; scref=(1.675e-04,1.237e-04,1.518e-04); uscref=(2.646e-05,2.908e-05,3.021e-05)

FieldCalibration MLE NO2

Pandora296s1_FieldCalibration_MLE_NO2.jpeg - 85 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora296s1_FieldCalibration_VCMLE_NO2.jpeg - 68 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(103); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=4.990e-03; wrmsmax=9.485e-04; atvarmax(%)=1.000e+02; amflim=(1.000,2.500); amflimvar=(0.000,1.000); perc=0.00; ngroup=5; timeaver(min)=0.00; redu=0; scref=(2.399e-05); uscref=(1.042e-05)

FieldCalibration MLE SO2

Pandora296s1_FieldCalibration_MLE_SO2.jpeg - 57 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora296s1_FieldCalibration_VCMLE_SO2.jpeg - 51 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=3.743e-04; wrmsmax=6.669e-04; atvarmax(%)=1.000e+02; amflim=(1.000,3.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=6; timeaver(min)=0.00; redu=2; scref=(9.783e-05); uscref=(4.541e-05)

FieldCalibration MLE HCHO

Pandora296s1_FieldCalibration_MLE_HCHO.jpeg - 53 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora296s1_FieldCalibration_VCMLE_HCHO.jpeg - 56 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
With Care

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes

SO2

Decision for SO2
Yes

HCHO

Decision for HCHO
Yes

H2O

Decision for H2O
Yes