Decision Summary

Upload-ready calibration protocol overview
Instrument P295s1
Session 3
Location TitoScaloPZ-CNR-IMAA
Calibration person Leonie Haunold
CF file Pandora295s1_CF_v3d20250801.txt
Software version 1.8.92
Calibration vector calV1
Reference day 20250913
Overall status development
O3Ext: With CareO3Synt: YesNO2: YesSO2: With CareHCHO: YesH2O: Yes
Instrument P295s1
Session 3
Events 315
Figures 17
Action time 5h 11m

Action Timeline

85 action intervals
Apr 23, 2026Apr 24, 2026May 11, 202607:27 AMWavelengthCalibrationCreateReferenceFieldCalibrationAssignCalFileWavelengthCalibration 9s finishedCreateReference 1m 7s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 37s finishedCreateReference 0s abortedCreateReference 1m 7s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 2m 19s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 22s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 33s abortedCreateReference 18s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 44s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 38s finishedCreateReference 0s finishedFieldCalibration 4m 35s abortedFieldCalibration 2m 9s finishedFieldCalibration 0s abortedFieldCalibration 54s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 2h 3m finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 16s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 16s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 2s abortedFieldCalibration 6s finishedFieldCalibration 0s finishedFieldCalibration 2h 46m finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s staleFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 9s abortedFieldCalibration 15s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 16s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 13s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 3s abortedAssignCalFile 2m 46s finishedWavelengthCalibration 0s open

Duration by Step

FieldCalibration4h 59mCreateReference8m 44sAssignCalFile2m 46sWavelengthCalibration9s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 2m 46s
CreateReference 55 27 8 8m 44s
FieldCalibration 240 54 10 4h 59m
No action 12 0 0 0s
WavelengthCalibration 5 1 0 9s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora295s1_WavelengthCalibration_SlitFuncParam.jpeg - 116 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora295s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 108 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora295s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 95 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora295s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 94 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora295s1_CreateReference_AODEST.jpeg - 49 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora295s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 82 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora295s1_CreateReference_DISPFIT(DISP).jpeg - 78 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora295s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 128 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora295s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 96 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora295s1_FieldCalibration_VC_O3.jpeg - 76 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora295s1_FieldCalibration_SCREFINT_O3.jpeg - 56 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=9.097e-05; wrmsmax=2.307e+00; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=50; timeaver(min)=0.00; redu=2; scref=(8.998e-05,7.776e-05,7.719e-05); uscref=(9.709e-06,3.599e-06,8.715e-06)

FieldCalibration MLE NO2

Pandora295s1_FieldCalibration_MLE_NO2.jpeg - 116 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora295s1_FieldCalibration_VCMLE_NO2.jpeg - 101 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=5.002e-03; wrmsmax=2.253e-01; atvarmax(%)=2.000e+01; amflim=(1.000,3.500); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(1.265e-05); uscref=(1.344e-05)

FieldCalibration MLE SO2

Pandora295s1_FieldCalibration_MLE_SO2.jpeg - 66 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora295s1_FieldCalibration_VCMLE_SO2.jpeg - 55 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=5.156e-03; wrmsmax=9.000e-04; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=2; scref=(1.662e-04); uscref=(9.198e-06)

FieldCalibration MLE HCHO

Pandora295s1_FieldCalibration_MLE_HCHO.jpeg - 69 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora295s1_FieldCalibration_VCMLE_HCHO.jpeg - 81 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
With Care
out2 flagged but not too bad

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
E-MLE used

SO2

Decision for SO2
With Care
zig-zag pattern

HCHO

Decision for HCHO
Yes
E-MLE used

H2O

Decision for H2O
Yes