Decision Summary

Upload-ready calibration protocol overview
Instrument P289s1
Session 31
Location -
Calibration person Hanna Rohringer
CF file Pandora289s1_CF_v31d20250921.txt
Software version 1.8.89
Calibration vector calV3
Reference day -
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P289s1
Session 31
Events 26
Figures 17
Action time 11m 22s

Action Timeline

8 action intervals
Jun 24, 202607:43 AM08:36 AMCreateReferenceFieldCalibrationAssignCalFileCreateReference 2s abortedFieldCalibration 1m 28s abortedFieldCalibration 1m 8s abortedFieldCalibration 28s abortedFieldCalibration 7m 41s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedAssignCalFile 36s finished

Duration by Step

FieldCalibration10m 45sAssignCalFile36sCreateReference2s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 36s
CreateReference 2 1 1 2s
FieldCalibration 12 6 4 10m 45s
No action 9 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora289s1_WavelengthCalibration_SlitFuncParam.jpeg - 168 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora289s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 143 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora289s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 123 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora289s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 177 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora289s1_CreateReference_AODEST.jpeg - 65 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora289s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 113 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora289s1_CreateReference_DISPFIT(DISP).jpeg - 108 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora289s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 197 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora289s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 122 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora289s1_FieldCalibration_VC_O3.jpeg - 104 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora289s1_FieldCalibration_SCREFINT_O3.jpeg - 77 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 redu=1 settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=4.636e-05; wrmsmax=9.000e-04; atvarmax(%)=2.000e+01; amflim=(1.000,4.500); amflimvar=(0.000,1.000); perc=0.00; ngroup=25; timeaver(min)=0.00; redu=1; scref=(1.712e-04,1.503e-04,1.600e-04); uscref=(1.213e-05,1.002e-05,1.039e-05)

FieldCalibration MLE NO2

Pandora289s1_FieldCalibration_MLE_NO2.jpeg - 187 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora289s1_FieldCalibration_VCMLE_NO2.jpeg - 91 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora289s1_FieldCalibration_MLE_SO2.jpeg - 67 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora289s1_FieldCalibration_VCMLE_SO2.jpeg - 54 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora289s1_FieldCalibration_MLE_HCHO.jpeg - 69 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora289s1_FieldCalibration_VCMLE_HCHO.jpeg - 48 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O