Decision Summary

Upload-ready calibration protocol overview
Instrument P26s1
Session 7
Location BoulderCO
Calibration person Manuel Gebetsberger
CF file Pandora26s1_CF_v1d20200523.txt
Software version 1.8.26
Calibration vector -
Reference day -
Overall status development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Instrument P26s1
Session 7
Events 95
Figures 1
Action time 1h 15m

Action Timeline

25 action intervals
Jul 01, 202011:41 AM09:10 PMCheckDispersionCreateReferenceAssignCalFileCheckDispersion 19m 34s finishedCheckDispersion 24s finishedCheckDispersion 0s finishedCheckDispersion 6m 47s finishedCheckDispersion 2m 45s finishedCreateReference 27m 10s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 5m 34s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 5s abortedCreateReference 2m 9s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 58s abortedCreateReference 7m 3s abortedCreateReference 2m 10s finishedCreateReference 0s finishedCreateReference 0s abortedAssignCalFile 39s finished

Duration by Step

CreateReference45m 9sCheckDispersion29m 30sAssignCalFile39s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 4 1 0 39s
CheckDispersion 17 5 0 29m 30s
CreateReference 66 19 7 45m 9s
No action 8 0 0 0s

Session Figures

1 figure from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora26s1_WavelengthCalibration_SlitFuncParam.jpeg - 135 KB
Slit function parameters

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2

SO2

Decision for SO2

HCHO

Decision for HCHO

H2O

Decision for H2O