Decision Summary
Upload-ready calibration protocol overview
Instrument
P245s1
Session
1
Location
-
Calibration person
-
CF file
-
Software version
1.8.49
Calibration vector
-
Reference day
-
Overall status
development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Action Timeline
No start/finish intervals found.No timeline intervals found
Duration by Step
No completed durations found
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| No actions loaded | ||||
Session Figures
7 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora245s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 105 KBCreate reference AODEST
Pandora245s1_CreateReference_AODEST.jpeg - 32 KBCreate reference SENSCORRFIN (SENS)
Pandora245s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 75 KBCreate reference WLCHANGE (DISP)
Pandora245s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 120 KBCreate reference SHIFT (CLOS)
Pandora245s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 68 KB
field calibration
O3
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Reference (P2s1) vs uncalibrated P245s1 slope of 1.022 and daily profiles matched ref instrument very well
O3Synt
Decision for O3Synt
NO2
Decision for NO2
Reference (P2s1) vs uncalibrated P245s1 slope of 1.01 and daily profiles almost identical compared with reference
SO2
Decision for SO2
Reference (P2s1) vs uncalibrated P245s1 slope of 1.396. SO2 calibration OK, but SO2 values low during clearance period, this will be improved with final field cal. Still given approval for clearance
HCHO
Decision for HCHO
Reference (P2s1) vs uncalibrated P245s1 slope of 1.023. Some similarities in shape but values not matching too great. Note: because most of these days were flagged out for HCHO in ref instrument also and other species looked good, this was still approved for clearance
H2O
Decision for H2O