Calibration Review

Upload-ready calibration protocol overview
Instrument P240s1
Session 11
Location Thessaloniki
Calibration person Ragi Rajagopalan
CF file Pandora240s1_CF_v11d20220916.txt
Software version p1-8
Calibration vector calV1
Reference day 20221030
Overall status development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: YesHCHO: With CareH2O: Not answered
Instrument P240s1
Session 11
Events 222
Figures 0
Action time 1h 0m

Action Timeline

61 action intervals
Aug 22, 202301:37 PM04:28 PMWavelengthCalibrationCreateReferenceFieldCalibrationAssignCalFileWavelengthCalibration 24s finishedCreateReference 7s abortedWavelengthCalibration 10s finishedCreateReference 52s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 12s finishedCreateReference 0s abortedCreateReference 59s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 44s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 27s abortedCreateReference 21s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 24s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 36s finishedCreateReference 0s finishedFieldCalibration 1m 54s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 44m 13s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 3m 39s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 1m 33s abortedFieldCalibration 2m 52s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 16s finishedFieldCalibration 5s aborted

Duration by Step

FieldCalibration54m 17sCreateReference5m 42sWavelengthCalibration34sAssignCalFile16s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 16s
CreateReference 47 24 8 5m 42s
FieldCalibration 155 34 2 54m 17s
No action 11 0 0 0s
WavelengthCalibration 6 2 0 34s

Session Figures

No session figures found for the loaded instrument/session.
No figures loaded

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
With Care
The W-shape O3-synt , O3-temp wrms, O3-Temp total columns has M shape some days .. seems to be associated with the W-shape in wrms

O3Synt

Decision for O3Synt
With Care

NO2

Decision for NO2
Yes
good overall

SO2

Decision for SO2
Yes
looks good

HCHO

Decision for HCHO
With Care
very large spread in both tropopsheric and total colums

H2O

Decision for H2O