Calibration Review
Upload-ready calibration protocol overview
Instrument
P240s1
Session
11
Location
Thessaloniki
Calibration person
Ragi Rajagopalan
CF file
Pandora240s1_CF_v11d20220916.txt
Software version
p1-8
Calibration vector
calV1
Reference day
20221030
Overall status
development
O3Ext: With CareO3Synt: With CareNO2: YesSO2: YesHCHO: With CareH2O: Not answered
Action Timeline
61 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AssignCalFile | 3 | 1 | 0 | 16s |
| CreateReference | 47 | 24 | 8 | 5m 42s |
| FieldCalibration | 155 | 34 | 2 | 54m 17s |
| No action | 11 | 0 | 0 | 0s |
| WavelengthCalibration | 6 | 2 | 0 | 34s |
Session Figures
No session figures found for the loaded instrument/session.No figures loaded
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
With Care
The W-shape O3-synt , O3-temp wrms, O3-Temp total columns has M shape some days .. seems to be associated with the W-shape in wrms
O3Synt
Decision for O3Synt
With Care
NO2
Decision for NO2
Yes
good overall
SO2
Decision for SO2
Yes
looks good
HCHO
Decision for HCHO
With Care
very large spread in both tropopsheric and total colums
H2O
Decision for H2O