Calibration Review
Upload-ready calibration protocol overview
Instrument
P240s1
Session
10
Location
LabGSFC
Calibration person
Ragi Rajagopalan
CF file
Pandora240s1_CF_v10d20221220.txt
Software version
p1-8
Calibration vector
-
Reference day
20230407
Overall status
development
O3Ext: YesO3Synt: YesNO2: YesSO2: YesHCHO: With CareH2O: Not answered
Action Timeline
98 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeBright | 12 | 1 | 0 | 49s |
| AssignCalFile | 6 | 2 | 0 | 46s |
| Convolution | 16 | 5 | 0 | 40s |
| CreateReference | 83 | 40 | 13 | 11m 54s |
| FieldCalibration | 199 | 38 | 3 | 1h 41m |
| NDFilters | 100 | 1 | 0 | 8m 28s |
| No action | 13 | 0 | 0 | 0s |
| SensitivityCalibration | 80 | 4 | 1 | 4m 40s |
| TempDependence | 7 | 1 | 0 | 29s |
| WavelengthCalibration | 22 | 6 | 1 | 8m 37s |
Session Figures
No session figures found for the loaded instrument/session.No figures loaded
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
No W-shape in wrms as the ref day is close to summer
O3Synt
Decision for O3Synt
Yes
NO2
Decision for NO2
Yes
good overall
SO2
Decision for SO2
Yes
looks good
HCHO
Decision for HCHO
With Care
HCHO is good in the summer as of now
H2O
Decision for H2O