Decision Summary
Upload-ready calibration protocol overview
Instrument
P237s1
Session
4
Location
AtlantaGA-SouthDeKalb
Calibration person
Stefanie Morhenn
CF file
Pandora237s1_CF_v4d20220528.txt
Software version
1.8.55
Calibration vector
-
Reference day
20230417
Overall status
development
O3Ext: YesO3Synt: YesNO2: YesSO2: With CareHCHO: With CareH2O: Yes
Action Timeline
85 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeBright | 2 | 1 | 1 | 14s |
| AnalyzeDark | 2 | 1 | 1 | 9s |
| AssignCalFile | 3 | 1 | 0 | 17s |
| Convolution | 2 | 1 | 1 | 1s |
| CreateReference | 74 | 36 | 11 | 27m 42s |
| FieldCalibration | 317 | 39 | 6 | 4h 23m |
| NDFilters | 2 | 1 | 1 | 6s |
| No action | 17 | 0 | 0 | 0s |
| SensitivityCalibration | 2 | 1 | 1 | 18s |
| TempDependence | 2 | 1 | 1 | 7s |
| WavelengthCalibration | 4 | 2 | 2 | 25s |
Session Figures
17 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora237s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 140 KBCreate reference SELSPECS
Pandora237s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 131 KBCreate reference SLERR (STRAY)
Pandora237s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 129 KBCreate reference AODEST
Pandora237s1_CreateReference_AODEST.jpeg - 57 KBCreate reference SENSCORRFIN (SENS)
Pandora237s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 115 KBCreate reference DISPFIT (DISP)
Pandora237s1_CreateReference_DISPFIT(DISP).jpeg - 108 KBCreate reference WLCHANGE (DISP)
Pandora237s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 114 KBCreate reference SHIFT (CLOS)
Pandora237s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 174 KB
field calibration
O3
field calibration
NO2
field calibration
SO2
field calibration
HCHO
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
Yes
interpolation used in AXC
O3Synt
Decision for O3Synt
Yes
NO2
Decision for NO2
Yes
MLE used, looks good
SO2
Decision for SO2
With Care
looks good
HCHO
Decision for HCHO
With Care
some days have double values in HCHO SC and wrms due to unknown hardware issues
H2O
Decision for H2O
Yes
looks good.