Decision Summary
Upload-ready calibration protocol overview
Instrument
P233s1
Session
1
Location
LabGSFC
Calibration person
Apoorva Pandey
CF file
Pandora233s1_CF_v1d20230902.txt
Software version
p1-8
Calibration vector
-
Reference day
20230902
Overall status
development
O3Ext: Not answeredO3Synt: Not answeredNO2: Not answeredSO2: Not answeredHCHO: Not answeredH2O: Not answered
Action Timeline
147 action intervalsDuration by Step
Action Summary
| Action | Events | Runs | Aborted | Time |
|---|---|---|---|---|
| AnalyzeBright | 15 | 1 | 0 | 7m 3s |
| AnalyzeDark | 17 | 2 | 0 | 3m 1s |
| AssignCalFile | 4 | 1 | 0 | 1m 7s |
| CheckFilters | 4 | 1 | 0 | 8s |
| CheckNoiseWave | 9 | 2 | 1 | 12s |
| Convolution | 16 | 5 | 0 | 4m 12s |
| CreateReference | 77 | 38 | 14 | 20m 13s |
| FieldCalibration | 125 | 44 | 7 | 2h 36m |
| NDFilters | 30 | 3 | 2 | 10m 20s |
| No action | 17 | 0 | 0 | 0s |
| SensitivityCalibration | 76 | 3 | 0 | 7m 47s |
| StrayLight | 52 | 16 | 6 | 6m 44s |
| TempDependence | 7 | 1 | 0 | 1m 48s |
| WavelengthCalibration | 106 | 29 | 12 | 7h 39m |
Session Figures
10 figures from Google Drive/figures
calprotocol section
Wavelength Calibration
reference period
Reference Day Decision Support
Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.
calprotocol section
Reference Period and CreateReference
Create reference L1CHARFILT
Pandora233s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 154 KBCreate reference SELSPECS
Pandora233s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 128 KBCreate reference SLERR (STRAY)
Pandora233s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 160 KBCreate reference AODEST
Pandora233s1_CreateReference_AODEST.jpeg - 53 KBCreate reference SENSCORRFIN (SENS)
Pandora233s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 122 KBCreate reference DISPFIT (DISP)
Pandora233s1_CreateReference_DISPFIT(DISP).jpeg - 94 KBCreate reference WLCHANGE (DISP)
Pandora233s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 187 KBCreate reference SHIFT (CLOS)
Pandora233s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 146 KB
field calibration
O3
Automatically harvested metadata
Automatically Harvested Metadata
Manual review and decision
Manual Review and Decision
Calibration notes, data-quality findings, and whole-period evaluation.Data Evaluation
Could there be an offset between VPs?
No
Are there breakpoint-analysis plots or suspicious discontinuities?
No
Do daily figures before and after the new VP look consistent?
No
Is there an AMF dependency for one or more VPs?
No
Is there seasonality, for example negative values in winter?
No
Does the daily shape follow the trop data roughly?
No
If there is a discrepancy, what are the DS and Sky viewing directions?
No
Is there a co-located or nearby instrument for comparison?
No
O3Ext
Decision for O3Ext
AXC looked good
O3Synt
Decision for O3Synt
NO2
Decision for NO2
ref method, looked good
SO2
Decision for SO2
ref method, looked good
HCHO
Decision for HCHO
ref method, looked good
H2O
Decision for H2O