Decision Summary

Upload-ready calibration protocol overview
Instrument P227s1
Session 4
Location NakhonPathom
Calibration person Leonie Haunold
CF file Pandora227s1_CF_v4d20250216.txt
Software version 1.8.92
Calibration vector calV1
Reference day 20260302-20260303
Overall status development
O3Ext: UnusableO3Synt: UnusableNO2: YesSO2: With CareHCHO: YesH2O: Yes
Instrument P227s1
Session 4
Events 310
Figures 17
Action time 59m 43s

Action Timeline

69 action intervals
Jun 16, 2026Jun 17, 2026Jun 18, 202609:46 AMAnalyzeDarkWavelengthCalibrationCreateReferenceFieldCalibrationTempDependenceAssignCalFileAnalyzeDark 1s abortedWavelengthCalibration 1m 8s abortedCreateReference 1m 35s abortedCreateReference 27s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 30s finishedCreateReference 0s abortedCreateReference 1m 48s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 2m 16s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 4s abortedCreateReference 24s finishedCreateReference 0s abortedCreateReference 18s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 2m 36s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 17s finishedCreateReference 0s finishedFieldCalibration 2m 17s abortedFieldCalibration 15m 50s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 1m 48s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedTempDependence 2s abortedFieldCalibration 12s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s staleFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 16m 38s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 8m 31s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 1m 0s finished

Duration by Step

FieldCalibration45m 16sCreateReference12m 16sWavelengthCalibration1m 8sAssignCalFile1m 0sTempDependence2sAnalyzeDark1s

Action Summary

Action Events Runs Aborted Time
AnalyzeDark 2 1 1 1s
AssignCalFile 3 1 0 1m 0s
CreateReference 58 26 9 12m 16s
FieldCalibration 230 38 2 45m 16s
No action 13 0 0 0s
TempDependence 2 1 1 2s
WavelengthCalibration 2 1 1 1m 8s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora227s1_WavelengthCalibration_SlitFuncParam.jpeg - 121 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora227s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 110 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora227s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 90 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora227s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 92 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora227s1_CreateReference_AODEST.jpeg - 48 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora227s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 86 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora227s1_CreateReference_DISPFIT(DISP).jpeg - 79 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora227s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 91 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora227s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 110 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora227s1_FieldCalibration_VC_O3.jpeg - 67 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora227s1_FieldCalibration_SCREFINT_O3.jpeg - 51 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.501e-04; wrmsmax=9.000e-04; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=100; timeaver(min)=0.00; redu=2; scref=(1.263e-04,1.036e-04,1.132e-04); uscref=(1.345e-05,3.551e-06,1.475e-05)

FieldCalibration MLE NO2

Pandora227s1_FieldCalibration_MLE_NO2.jpeg - 143 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora227s1_FieldCalibration_VCMLE_NO2.jpeg - 100 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=9.976e-03; wrmsmax=2.272e-02; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(3.937e-05); uscref=(4.125e-06)

FieldCalibration MLE SO2

Pandora227s1_FieldCalibration_MLE_SO2.jpeg - 63 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora227s1_FieldCalibration_VCMLE_SO2.jpeg - 55 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=3.179e-04; wrmsmax=3.544e-03; atvarmax(%)=2.000e+01; amflim=(1.000,2.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=2; scref=(2.912e-04); uscref=(5.250e-05)

FieldCalibration MLE HCHO

Pandora227s1_FieldCalibration_MLE_HCHO.jpeg - 65 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora227s1_FieldCalibration_VCMLE_HCHO.jpeg - 90 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Unusable
out2 inverse-u and flagged, ous1 u-shaped

O3Synt

Decision for O3Synt
Unusable

NO2

Decision for NO2
Yes

SO2

Decision for SO2
With Care
a bit wiggly but all <1e-4

HCHO

Decision for HCHO
Yes

H2O

Decision for H2O
Yes