Decision Summary

Upload-ready calibration protocol overview
Instrument P226s1
Session 2
Location GreenbeltMD
Calibration person Bryan Place
CF file Pandora226s1_CF_v2d20240301.txt
Software version p1-8
Calibration vector -
Reference day 20240324
Overall status development
O3Ext: YesO3Synt: YesNO2: YesSO2: YesHCHO: YesH2O: Not answered
Instrument P226s1
Session 2
Events 207
Figures 17
Action time 3h 6m

Action Timeline

65 action intervals
Apr 03, 2024Apr 04, 2024Jan 07, 202510:45 PMCreateReferenceSensitivityCalibrationFieldCalibrationAssignCalFileCreateReference 1m 8s finishedCreateReference 0s abortedCreateReference 55s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 28s finishedCreateReference 0s abortedCreateReference 11s abortedSensitivityCalibration 11s abortedCreateReference 1m 1s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 45s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 32s abortedCreateReference 20s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 26s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 2m 13s finishedCreateReference 0s finishedFieldCalibration 0s staleFieldCalibration 26s abortedFieldCalibration 0s staleFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 2h 4m finishedFieldCalibration 14m 8s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 16m 7s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 22m 18s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedCreateReference 7s abortedAssignCalFile 23s finished

Duration by Step

FieldCalibration2h 57mCreateReference9m 7sAssignCalFile23sSensitivityCalibration11s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 23s
CreateReference 53 27 10 9m 7s
FieldCalibration 122 34 1 2h 57m
No action 21 0 0 0s
SensitivityCalibration 8 1 1 11s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora226s1_WavelengthCalibration_SlitFuncParam.jpeg - 252 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora226s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 123 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora226s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 102 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora226s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 89 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora226s1_CreateReference_AODEST.jpeg - 38 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora226s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 94 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora226s1_CreateReference_DISPFIT(DISP).jpeg - 70 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora226s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 132 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora226s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 109 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora226s1_FieldCalibration_VC_O3.jpeg - 73 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora226s1_FieldCalibration_SCREFINT_O3.jpeg - 55 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora226s1_FieldCalibration_MLE_NO2.jpeg - 99 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora226s1_FieldCalibration_VCMLE_NO2.jpeg - 66 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora226s1_FieldCalibration_MLE_SO2.jpeg - 54 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora226s1_FieldCalibration_VCMLE_SO2.jpeg - 53 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora226s1_FieldCalibration_MLE_HCHO.jpeg - 55 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora226s1_FieldCalibration_VCMLE_HCHO.jpeg - 58 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
Both O3 code retrievals looked good + out2 wrms was below threshold

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
REF method (P2). NO2 columns + wrms shape look good and match reference

SO2

Decision for SO2
Yes
REF method (P2). SO2 columns + wrms shape look good and match reference

HCHO

Decision for HCHO
Yes
REF method (P2). HCHO columns + wrms shape looked OK and matched reference most days

H2O

Decision for H2O