Decision Summary

Upload-ready calibration protocol overview
Instrument P222s1
Session 6
Location -
Calibration person Stefanie Morhenn
CF file Pandora222s1_CF_v6d20250923.txt
Software version 1.8.88
Calibration vector CalV3
Reference day -
Overall status development
O3Ext: UnusableO3Synt: UnusableNO2: With CareSO2: YesHCHO: With CareH2O: Yes
Instrument P222s1
Session 6
Events 25
Figures 17
Action time 6m 22s

Action Timeline

8 action intervals
Apr 30, 202608:27 AM10:14 AMFieldCalibrationAssignCalFileFieldCalibration 4m 45s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 1m 15s abortedAssignCalFile 0s openAssignCalFile 22s finished

Duration by Step

FieldCalibration4m 45sAssignCalFile1m 37s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 7 2 1 1m 37s
FieldCalibration 8 5 0 4m 45s
No action 10 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora222s1_WavelengthCalibration_SlitFuncParam.jpeg - 153 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora222s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 150 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora222s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 123 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora222s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 135 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora222s1_CreateReference_AODEST.jpeg - 64 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora222s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 121 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora222s1_CreateReference_DISPFIT(DISP).jpeg - 116 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora222s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 110 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora222s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 190 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora222s1_FieldCalibration_VC_O3.jpeg - 96 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora222s1_FieldCalibration_SCREFINT_O3.jpeg - 73 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora222s1_FieldCalibration_MLE_NO2.jpeg - 151 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora222s1_FieldCalibration_VCMLE_NO2.jpeg - 104 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora222s1_FieldCalibration_MLE_SO2.jpeg - 99 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora222s1_FieldCalibration_VCMLE_SO2.jpeg - 90 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO redu=1 settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=4.753e-04; wrmsmax=5.816e-03; atvarmax(%)=5.000e+01; amflim=(1.020,5.000); amflimvar=(0.000,1.000); perc=10.00; ngroup=10; timeaver(min)=0.00; redu=1; scref=(2.905e-04); uscref=(3.133e-05)

FieldCalibration MLE HCHO

Pandora222s1_FieldCalibration_MLE_HCHO.jpeg - 102 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora222s1_FieldCalibration_VCMLE_HCHO.jpeg - 107 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Unusable
inverse-u and flagged out, unchanged from ooB

O3Synt

Decision for O3Synt
Unusable

NO2

Decision for NO2
With Care
partially flagged out

SO2

Decision for SO2
Yes
looks good

HCHO

Decision for HCHO
With Care
partially to high - flagged out manually

H2O

Decision for H2O
Yes
looks good