Decision Summary

Upload-ready calibration protocol overview
Instrument P217s1
Session 7
Location -
Calibration person Manuel Roca
CF file Pandora217s1_CF_v7d20220418.txt
Software version 1.8.49
Calibration vector calV1
Reference day -
Overall status development
O3Ext: YesO3Synt: YesNO2: YesSO2: YesHCHO: With CareH2O: Yes
Instrument P217s1
Session 7
Events 17
Figures 17
Action time 10m 27s

Action Timeline

6 action intervals
Mar 13, 202312:20 PM12:57 PMFieldCalibrationAssignCalFileFieldCalibration 8m 37s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedAssignCalFile 1m 50s finished

Duration by Step

FieldCalibration8m 37sAssignCalFile1m 50s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 1m 50s
FieldCalibration 8 5 0 8m 37s
No action 6 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora217s1_WavelengthCalibration_SlitFuncParam.jpeg - 227 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora217s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 202 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora217s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 190 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora217s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 195 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora217s1_CreateReference_AODEST.jpeg - 100 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora217s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 168 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora217s1_CreateReference_DISPFIT(DISP).jpeg - 144 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora217s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 288 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora217s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 190 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora217s1_FieldCalibration_VC_O3.jpeg - 164 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora217s1_FieldCalibration_SCREFINT_O3.jpeg - 111 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora217s1_FieldCalibration_MLE_NO2.jpeg - 273 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora217s1_FieldCalibration_VCMLE_NO2.jpeg - 248 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora217s1_FieldCalibration_MLE_SO2.jpeg - 146 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora217s1_FieldCalibration_VCMLE_SO2.jpeg - 143 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO redu=1 settings applied fw1out=(); fw2out=(); dqf2in=(103); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=7.638e-04; wrmsmax=8.119e-03; atvarmax(%)=4.000e+01; amflim=(1.081,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=1; scref=(1.006e-04); uscref=(3.812e-05)

FieldCalibration MLE HCHO

Pandora217s1_FieldCalibration_MLE_HCHO.jpeg - 150 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora217s1_FieldCalibration_VCMLE_HCHO.jpeg - 220 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes

SO2

Decision for SO2
Yes

HCHO

Decision for HCHO
With Care
Presents negative values first months of 2023. However it could be seasonal

H2O

Decision for H2O
Yes