Decision Summary

Upload-ready calibration protocol overview
Instrument P213s1
Session 5
Location LabGSFC
Calibration person Bryan Place
CF file Pandora213s1_CF_v5d20250829.txt
Software version 1.8.73
Calibration vector -
Reference day 20251009
Overall status development
O3Ext: YesO3Synt: YesNO2: YesSO2: YesHCHO: YesH2O: Not answered
Instrument P213s1
Session 5
Events 419
Figures 17
Action time 3h 27m

Action Timeline

90 action intervals
Oct 28, 2025Oct 29, 2025Dec 30, 202504:15 PMCheckFiltersCheckNoiseWaveAnalyzeDarkAnalyzeBrightWavelengthCalibrationTempDependenceNDFiltersStrayLightSensitivityCalibrationConvolutionCreateReferenceFieldCalibrationAssignCalFileCheckFilters 49s finishedCheckNoiseWave 17s finishedAnalyzeDark 2m 32s finishedAnalyzeBright 4m 52s abortedAnalyzeBright 1m 21s finishedWavelengthCalibration 2m 9s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedWavelengthCalibration 0s finishedTempDependence 46s finishedNDFilters 1m 17s finishedStrayLight 47s finishedStrayLight 0s finishedStrayLight 0s abortedSensitivityCalibration 4m 25s finishedSensitivityCalibration 0s finishedSensitivityCalibration 0s finishedConvolution 45s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedConvolution 0s finishedCreateReference 9s abortedCreateReference 32s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 20s finishedCreateReference 0s abortedCreateReference 1m 24s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 49m 44s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 34s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 20s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 21s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 1m 15s finishedCreateReference 0s finishedFieldCalibration 12m 41s finishedFieldCalibration 4m 32s finishedFieldCalibration 0s abortedFieldCalibration 21s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 1h 2m finishedFieldCalibration 0s abortedFieldCalibration 4m 36s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s abortedFieldCalibration 19m 44s abortedFieldCalibration 14m 56s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 6m 5s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 5m 48s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedCreateReference 6s abortedFieldCalibration 28s abortedAnalyzeBright 2s abortedAssignCalFile 29s finished

Duration by Step

FieldCalibration2h 11mCreateReference54m 44sAnalyzeBright6m 15sSensitivityCalibration4m 25sAnalyzeDark2m 32sWavelengthCalibration2m 9sNDFilters1m 17sCheckFilters49sStrayLight47sTempDependence46sConvolution45sAssignCalFile29s

Action Summary

Action Events Runs Aborted Time
AnalyzeBright 18 3 2 6m 15s
AnalyzeDark 10 1 0 2m 32s
AssignCalFile 4 1 0 29s
CheckFilters 4 1 0 49s
CheckNoiseWave 6 1 0 17s
Convolution 16 5 0 45s
CreateReference 56 28 9 54m 44s
FieldCalibration 159 37 5 2h 11m
NDFilters 16 1 0 1m 17s
No action 18 0 0 0s
SensitivityCalibration 64 3 0 4m 25s
StrayLight 22 3 1 47s
TempDependence 7 1 0 46s
WavelengthCalibration 19 5 0 2m 9s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora213s1_WavelengthCalibration_SlitFuncParam.jpeg - 126 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora213s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 115 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora213s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 84 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora213s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 83 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora213s1_CreateReference_AODEST.jpeg - 38 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora213s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 79 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora213s1_CreateReference_DISPFIT(DISP).jpeg - 72 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora213s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 88 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora213s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 96 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora213s1_FieldCalibration_VC_O3.jpeg - 75 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora213s1_FieldCalibration_SCREFINT_O3.jpeg - 54 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=8.350e-04; wrmsmax=3.289e+00; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=40; timeaver(min)=0.00; redu=2; scref=(1.951e-04,1.680e-04,1.787e-04); uscref=(1.222e-05,6.530e-06,1.440e-05)

FieldCalibration MLE NO2

Pandora213s1_FieldCalibration_MLE_NO2.jpeg - 122 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora213s1_FieldCalibration_VCMLE_NO2.jpeg - 72 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=1.313e-02; wrmsmax=1.132e-01; atvarmax(%)=2.000e+01; amflim=(1.000,3.000); amflimvar=(0.000,1.000); perc=0.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(2.390e-05); uscref=(1.400e-05)

FieldCalibration MLE SO2

Pandora213s1_FieldCalibration_MLE_SO2.jpeg - 62 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora213s1_FieldCalibration_VCMLE_SO2.jpeg - 60 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=6.151e-03; wrmsmax=4.352e-02; atvarmax(%)=2.000e+01; amflim=(1.000,5.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=2; scref=(1.304e-04); uscref=(2.153e-05)

FieldCalibration MLE HCHO

Pandora213s1_FieldCalibration_MLE_HCHO.jpeg - 61 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora213s1_FieldCalibration_VCMLE_HCHO.jpeg - 72 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Yes
AXC-T estimated O3T: 230K, clim:227K. All O3 products looked good

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
EMLE used, NO2 columns and wrms looked good + matched REF instrument

SO2

Decision for SO2
Yes
EMLE used, SO2 columns looked good and less noisy than REF instrument

HCHO

Decision for HCHO
Yes
EMLE used, HCHO columns looked good (less noisy than REF) and matched REF columns shape well

H2O

Decision for H2O