Decision Summary

Upload-ready calibration protocol overview
Instrument P212s1
Session 6
Location -
Calibration person Hanna Rohringer
CF file Pandora212s1_CF_v6d20240309.txt
Software version 1.8.60
Calibration vector -
Reference day -
Overall status development
O3Ext: With CareO3Synt: YesNO2: YesSO2: With CareHCHO: With CareH2O: Yes
Instrument P212s1
Session 6
Events 63
Figures 17
Action time 1h 9m

Action Timeline

24 action intervals
Sep 26, 2024Oct 04, 2024Oct 13, 2024Oct 14, 202401:16 PM08:33 AMFieldCalibrationAssignCalFileCreateReferenceFieldCalibration 21m 13s finishedFieldCalibration 0s abortedFieldCalibration 12s abortedFieldCalibration 5m 21s finishedFieldCalibration 0s abortedAssignCalFile 57s finishedFieldCalibration 53s abortedFieldCalibration 22s abortedFieldCalibration 0s staleAssignCalFile 5s abortedCreateReference 4s abortedFieldCalibration 28s finishedFieldCalibration 0s finishedAssignCalFile 1m 8s finishedFieldCalibration 16m 40s abortedFieldCalibration 20m 48s finishedFieldCalibration 0s abortedFieldCalibration 18s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 28s abortedAssignCalFile 53s finished

Duration by Step

FieldCalibration1h 6mAssignCalFile3m 2sCreateReference4s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 11 4 1 3m 2s
CreateReference 2 1 1 4s
FieldCalibration 36 18 8 1h 6m
No action 14 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora212s1_WavelengthCalibration_SlitFuncParam.jpeg - 155 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora212s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 169 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora212s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 136 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora212s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 132 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora212s1_CreateReference_AODEST.jpeg - 57 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora212s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 122 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora212s1_CreateReference_DISPFIT(DISP).jpeg - 98 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora212s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 195 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora212s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 131 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora212s1_FieldCalibration_VC_O3.jpeg - 103 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora212s1_FieldCalibration_SCREFINT_O3.jpeg - 78 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora212s1_FieldCalibration_MLE_NO2.jpeg - 113 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora212s1_FieldCalibration_VCMLE_NO2.jpeg - 77 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(103); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=9.113e-04; wrmsmax=6.824e-02; atvarmax(%)=8.000e+01; amflim=(1.000,2.800); amflimvar=(0.000,1.000); perc=20.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(5.109e-05); uscref=(1.204e-05)

FieldCalibration MLE SO2

Pandora212s1_FieldCalibration_MLE_SO2.jpeg - 72 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora212s1_FieldCalibration_VCMLE_SO2.jpeg - 55 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO MLE settings applied fw1out=(); fw2out=(); dqf2in=(103); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=9.387e-01; wrmsmax=2.457e-01; atvarmax(%)=8.000e+01; amflim=(1.000,2.200); amflimvar=(0.000,1.000); perc=0.00; ngroup=8; timeaver(min)=0.00; redu=0; scref=(2.015e-04); uscref=(2.979e-04)

FieldCalibration MLE HCHO

Pandora212s1_FieldCalibration_MLE_HCHO.jpeg - 79 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora212s1_FieldCalibration_VCMLE_HCHO.jpeg - 73 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
With Care
same as 5

O3Synt

Decision for O3Synt
Yes

NO2

Decision for NO2
Yes
same as 5

SO2

Decision for SO2
With Care
winter needs to be flagged

HCHO

Decision for HCHO
With Care
some days need to be falgged

H2O

Decision for H2O
Yes
ok