Decision Summary

Upload-ready calibration protocol overview
Instrument P212s1
Session 5
Location Pontianak
Calibration person Hanna Rohringer
CF file Pandora212s1_CF_v5d20240309.txt
Software version p1-8
Calibration vector -
Reference day 20240505
Overall status development
O3Ext: With CareO3Synt: Not answeredNO2: YesSO2: UnusableHCHO: UnusableH2O: Yes
Instrument P212s1
Session 5
Events 418
Figures 17
Action time 1h 27m

Action Timeline

93 action intervals
Sep 05, 2024Sep 06, 2024Sep 09, 2024Sep 12, 2024Sep 24, 202401:35 PMWavelengthCalibrationSensitivityCalibrationCreateReferenceFieldCalibrationAssignCalFileWavelengthCalibration 10s finishedSensitivityCalibration 19s finishedSensitivityCalibration 0s finishedCreateReference 32s finishedCreateReference 0s abortedCreateReference 16s finishedCreateReference 0s abortedCreateReference 13s abortedCreateReference 34s finishedCreateReference 0s abortedCreateReference 33s finishedCreateReference 0s abortedCreateReference 19s abortedCreateReference 3s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 39s finishedCreateReference 0s abortedCreateReference 2m 14s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 2m 48s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 23s abortedCreateReference 8s abortedCreateReference 28s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 2m 30s finishedCreateReference 0s finishedCreateReference 0s finishedCreateReference 0s abortedCreateReference 36s finishedCreateReference 0s finishedCreateReference 1s abortedFieldCalibration 2m 21s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s staleFieldCalibration 19s abortedFieldCalibration 4m 18s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 11s abortedFieldCalibration 58s abortedFieldCalibration 19m 17s finishedFieldCalibration 0s abortedFieldCalibration 1m 8s abortedFieldCalibration 55s finishedFieldCalibration 2m 50s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 9m 37s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 7m 48s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 0s finishedFieldCalibration 3s abortedFieldCalibration 4m 11s abortedFieldCalibration 4m 59s abortedFieldCalibration 13s abortedFieldCalibration 4m 37s abortedFieldCalibration 52s abortedFieldCalibration 1m 11s abortedFieldCalibration 4m 37s abortedAssignCalFile 2m 51s finishedFieldCalibration 1m 51s aborted

Duration by Step

FieldCalibration1h 12mCreateReference12m 16sAssignCalFile2m 51sSensitivityCalibration19sWavelengthCalibration10s

Action Summary

Action Events Runs Aborted Time
AssignCalFile 3 1 0 2m 51s
CreateReference 72 36 15 12m 16s
FieldCalibration 285 52 14 1h 12m
No action 19 0 0 0s
SensitivityCalibration 36 2 0 19s
WavelengthCalibration 3 1 0 10s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora212s1_WavelengthCalibration_SlitFuncParam.jpeg - 155 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora212s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 169 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora212s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 136 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora212s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 132 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora212s1_CreateReference_AODEST.jpeg - 57 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora212s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 122 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora212s1_CreateReference_DISPFIT(DISP).jpeg - 98 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora212s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 195 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora212s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 131 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora212s1_FieldCalibration_VC_O3.jpeg - 103 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora212s1_FieldCalibration_SCREFINT_O3.jpeg - 78 KB
FieldCalibration SCREFINT O3
field calibration

NO2

NO2 E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=2.352e-04; wrmsmax=2.329e+00; atvarmax(%)=4.000e+01; amflim=(1.190,4.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=40; timeaver(min)=0.00; redu=2; scref=(7.712e-05,5.701e-05,6.095e-05); uscref=(2.840e-05,1.983e-05,2.909e-05)

FieldCalibration MLE NO2

Pandora212s1_FieldCalibration_MLE_NO2.jpeg - 113 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora212s1_FieldCalibration_VCMLE_NO2.jpeg - 77 KB
FieldCalibration VCMLE NO2
field calibration

SO2

SO2 MLE settings applied fw1out=(); fw2out=(); dqf2in=(103); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=3.113e-03; wrmsmax=6.824e-02; atvarmax(%)=8.000e+01; amflim=(1.000,1.800); amflimvar=(0.000,1.000); perc=2.00; ngroup=10; timeaver(min)=0.00; redu=0; scref=(4.825e-05); uscref=(1.582e-05)

FieldCalibration MLE SO2

Pandora212s1_FieldCalibration_MLE_SO2.jpeg - 67 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora212s1_FieldCalibration_VCMLE_SO2.jpeg - 52 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

HCHO E-MLE settings applied fw1out=(); fw2out=(); dqf2in=(103); ddlim=(-1.000,-1.000); shilim(nm)=(-0.050,0.050); uscmax=2.387e-04; wrmsmax=2.457e-02; atvarmax(%)=6.000e+01; amflim=(1.050,3.000); amflimvar=(0.000,1.000); perc=50.00; ngroup=10; timeaver(min)=0.00; redu=2; scref=(4.031e-04); uscref=(1.144e-04)

FieldCalibration MLE HCHO

Pandora212s1_FieldCalibration_MLE_HCHO.jpeg - 83 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora212s1_FieldCalibration_VCMLE_HCHO.jpeg - 73 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
With Care
bellyShaped O3Ext

O3Synt

Decision for O3Synt

NO2

Decision for NO2
Yes
ok

SO2

Decision for SO2
Unusable
underestimated

HCHO

Decision for HCHO
Unusable
overestimated

H2O

Decision for H2O
Yes
ok