Decision Summary

Upload-ready calibration protocol overview
Instrument P20s1
Session 9
Location Busan
Calibration person Hanna Rohringer
CF file C:/Blick-dev18/data/calresults/Pandora20s1/session9/Pandora20s1_CF_session9.txt
Software version p1-8
Calibration vector -
Reference day -
Overall status development
O3Ext: UnusableO3Synt: UnusableNO2: With CareSO2: UnusableHCHO: UnusableH2O: Yes
Instrument P20s1
Session 9
Events 38
Figures 17
Action time 52m 43s

Action Timeline

3 action intervals
May 22, 202609:17 AM10:26 AMFieldCalibrationFieldCalibration 52m 43s finishedFieldCalibration 0s finishedFieldCalibration 0s finished

Duration by Step

FieldCalibration52m 43s

Action Summary

Action Events Runs Aborted Time
FieldCalibration 30 3 0 52m 43s
No action 8 0 0 0s

Session Figures

17 figures from Google Drive/figures
calprotocol section

Wavelength Calibration

Slit function parameters

Pandora20s1_WavelengthCalibration_SlitFuncParam.jpeg - 118 KB
Slit function parameters
reference period

Reference Day Decision Support

Upload the FOV and L0 raw-count evidence used to justify the selected reference day before reviewing the CreateReference figures.

calprotocol section

Reference Period and CreateReference

Create reference L1CHARFILT

Pandora20s1_CreateReference_L1CHARFILT_OPEN_U340.jpeg - 171 KB
Create reference L1CHARFILT

Create reference SELSPECS

Pandora20s1_CreateReference_SELSPECS_OPEN_U340.jpeg - 117 KB
Create reference SELSPECS

Create reference SLERR (STRAY)

Pandora20s1_CreateReference_SLERR(STRAY)_OPEN_U340.jpeg - 142 KB
Create reference SLERR (STRAY)

Create reference AODEST

Pandora20s1_CreateReference_AODEST.jpeg - 63 KB
Create reference AODEST

Create reference SENSCORRFIN (SENS)

Pandora20s1_CreateReference_SENSCORRFIN(SENS)_OPEN_U340.jpeg - 123 KB
Create reference SENSCORRFIN (SENS)

Create reference DISPFIT (DISP)

Pandora20s1_CreateReference_DISPFIT(DISP).jpeg - 104 KB
Create reference DISPFIT (DISP)

Create reference WLCHANGE (DISP)

Pandora20s1_CreateReference_WLCHANGE(DISP)_OPEN_U340.jpeg - 194 KB
Create reference WLCHANGE (DISP)

Create reference SHIFT (CLOS)

Pandora20s1_CreateReference_SHIFT(CLOS)_OPEN_U340.jpeg - 179 KB
Create reference SHIFT (CLOS)
field calibration

O3

FieldCalibration VC O3

Pandora20s1_FieldCalibration_VC_O3.jpeg - 60 KB
FieldCalibration VC O3

FieldCalibration SCREFINT O3

Pandora20s1_FieldCalibration_SCREFINT_O3.jpeg - 43 KB
FieldCalibration SCREFINT O3
field calibration

NO2

FieldCalibration MLE NO2

Pandora20s1_FieldCalibration_MLE_NO2.jpeg - 171 KB
FieldCalibration MLE NO2

FieldCalibration VCMLE NO2

Pandora20s1_FieldCalibration_VCMLE_NO2.jpeg - 55 KB
FieldCalibration VCMLE NO2
field calibration

SO2

FieldCalibration MLE SO2

Pandora20s1_FieldCalibration_MLE_SO2.jpeg - 49 KB
FieldCalibration MLE SO2

FieldCalibration VCMLE SO2

Pandora20s1_FieldCalibration_VCMLE_SO2.jpeg - 40 KB
FieldCalibration VCMLE SO2
field calibration

HCHO

FieldCalibration MLE HCHO

Pandora20s1_FieldCalibration_MLE_HCHO.jpeg - 49 KB
FieldCalibration MLE HCHO

FieldCalibration VCMLE HCHO

Pandora20s1_FieldCalibration_VCMLE_HCHO.jpeg - 48 KB
FieldCalibration VCMLE HCHO

Automatically harvested metadata

Automatically Harvested Metadata

Manual review and decision

Manual Review and Decision

Calibration notes, data-quality findings, and whole-period evaluation.

Data Evaluation

Could there be an offset between VPs?

No

Are there breakpoint-analysis plots or suspicious discontinuities?

No

Do daily figures before and after the new VP look consistent?

No

Is there an AMF dependency for one or more VPs?

No

Is there seasonality, for example negative values in winter?

No

Does the daily shape follow the trop data roughly?

No

If there is a discrepancy, what are the DS and Sky viewing directions?

No

Is there a co-located or nearby instrument for comparison?

No

O3Ext

Decision for O3Ext
Unusable
high res change all u340 porducts not usable - not calibrated

O3Synt

Decision for O3Synt
Unusable

NO2

Decision for NO2
With Care
some period with high res change are flagged

SO2

Decision for SO2
Unusable
high res change all u340 porducts not usable - not calibrated

HCHO

Decision for HCHO
Unusable
high res change all u340 porducts not usable - not calibrated

H2O

Decision for H2O
Yes
ok